6.12. SETUP MORE VARS: Internal Variables (VARS)
The M360E has several-user adjustable software variables, which define certain operational parameters.
Usually, these variables are automatically set by the instrument’s firmware, but can be manually re-defined using
the VARS menu. Table 6-14 lists all variables that are available within the 818 password protected level.
Table 6-14: Variable Names (VARS) Revision B.3
NO. VARIABLE DESCRIPTION ALLOWED
VALUES
0 DAS_HOLD_OFF
Changes the internal data acquisition system (iDAS) hold-off
time, which is the duration when data are not stored in the
iDAS because the software considers the data to be
questionable. That is the case during warm-up or just after
the instrument returns from one of its calibration modes to
SAMPLE mode. DAS_HOLD_OFF can be disabled entirely
in each iDAS channel.
Can be between 0.5
and 20 minutes
Default=15 min.
1 CONC_PRECISION
Allows the user to set the number of significant digits to the
right of the decimal point display of concentration and
stability values.
AUTO, 1, 2, 3, 4
Default=AUTO
2 STABIL_GAS Selects which gas measurement is displayed when the STABIL
test function is selected. CO; CO2 & O21
3 DYN_ZERO
Dynamic zero automatically adjusts offset and slope of the
CO2 response when performing a zero point calibration
during an AutoCal (Chapter 7).
ON/OFF
4 DYN_SPAN
Dynamic span automatically adjusts slope and slope of the
CO2 response when performing a zero point calibration
during an AutoCal (Chapter 7).
Note that the DYN_ZERO and DYN_SPAN features are not
allowed for applications requiring EPA equivalency.
ON/OFF
5 CLOCK_ADJ
Adjusts the speed of the analyzer’s clock. Choose the +
sign if the clock is too slow, choose the - sign if the clock is
too fast.
-60 to +60 s/day
1 O2 option is only available in analyzers with o2 sensor options installed.