Model 360E Instruction Manual Operating Instructions

6.13.9. Test Channel Output

When activated, output channel A4 can be used to report one of the test functions viewable from the SAMPLE

mode display. To activate the A4 channel and select a test function, follow this key sequence:

SETUP X.X ENTER DIAG PASS: 818
8 1 8 ENTR EXIT
SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX
< TST TST > CAL SETUP
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
DIAG SIGNAL I / O
NEXT ENTR EXIT
DIAG ANALOG OUTPUT
PREV NEXT ENTR EXIT
DIAG TCHN TEST CHANNEL: NONE
NEXT ENTR EXIT
DIAG TEST CHAN OUTPUT
PREV NEXT ENTR EXIT
Continue to press NEXT until …
DIAG TCHN TEST CHANNEL: CO2 MEASURE
PREV NEXT ENTR EXIT
PressPREV or NEXT
to move through the
list of available
parameters
(Table 6-13)
Press EXIT to
return to the
DIAG menu
Press ENTR to select
the displayed
parameter activating
the test channel.
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG ALRM EXIT

Table 6-22: Test Parameters Available for Analog Output A4

TEST CHANNEL ZERO FULL SCALE

NONE Test Channel is turned off

CO2 MEASURE 0 mV 5000 mV*

CO2 REFERENCE 0 mV 5000 mV*

SAMPLE PRESS 0 "Hg 40 "Hg
SAMPLE FLOW 0 cc/m 1000 cc/m
SAMPLE TEMP 0C 70C
BENCH TEMP 0C 70C
WHEEL TEMP 0C 70C
CHASSIS TEMP 0C 70C
PHT DRIVE 0 mV 5000 mV
* This refers to the internal voltage level of the function NOT the output signal level
of the Test channel itself.