Model 360E Instruction Manual APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4
SETUP VARIABLE NUMERIC
UNITS DEFAULT
VALUE VALUE RANGE DESCRIPTION
BAUD_RATE — 19200
1
300, 1200, 2400,
4800, 9600,
19200, 38400,
57600, 115200
RS-232 COM1 baud rate.
MODEM_INIT “AT Y0 &D0 &H0
&I0 S0=2 &B0 &N6
&M0 E0 Q1 &W0” 1
Any character in the
allowed character set.
Up to 100 characters
long.
RS-232 COM1 modem initialization
string. Sent verbatim plus carriage
return to modem on power up or
manually.
RS232_MODE2 BitFlag 0 0–65535
RS-232 COM2 mode flags.
(Same settings as RS232_MODE.)
BAUD_RATE2 — 19200
1
300, 1200,
2400, 4800,
9600, 19200,
38400, 57600,
115200
RS-232 COM2 baud rate.
MODEM_INIT2 “AT Y0 &D0 &H0
&I0 S0=2 &B0 &N6
&M0 E0 Q1 &W0” 1
Any character in the
allowed character set.
Up to 100 characters
long.
RS-232 COM2 modem initialization
string. Sent verbatim plus carriage
return to modem on power up or
manually.
RS232_PASS Password 940331 0–999999 RS-232 log on password.
MACHINE_ID ID 320 0–9999 Unique ID number for instrument.
COMMAND_PROMPT — “Cmd>
1
Any character in the
allowed character set.
Up to 100 characters
long.
RS-232 interface command prompt.
Displayed only if enabled with
RS232_MODE variable.
TEST_CHAN_ID — NONE
1
NONE,
CO MEASURE,
CO REFERENCE,
VACUUM PRESSURE,
SAMPLE PRESSURE,
SAMPLE FLOW,
SAMPLE TEMP,
BENCH TEMP,
WHEEL TEMP,
O2 CELL TEMP 5,
CHASSIS TEMP,
PHT DRIVE
Diagnostic analog output ID.
REMOTE_CAL_MODE — LOW
1 LOW,
HIGH
Range to calibrate during contact
closure or Hessen calibration.
PASS_ENABLE — OFF ON, OFF
ON enables passwords; OFF
disables them.
STABIL_GAS — CO
1 CO,
O2 5
Selects gas for stability
measurement.
STABIL_FREQ Seconds 10 1–300
Stability measurement sampling
frequency.
STABIL_SAMPLES Samples 25 2–40
Number of samples in concentration
stability reading.
PHOTO_TEMP_SET mV
2500
Warnings:
250–4750
0–5000 Photometer temperature warning
limits. Set point is not used.
05233 Rev G.4 A-13