Model 360E Instruction Manual Operating Instructions

To enter the signal I/O test mode, press:

SAMPLE RANGE = 500.000 PPM CO2 =X.XXX
< TST TST > CAL SETUP
EXAMPLE
SAMPLE ENTER SETUP PASS: 818
8 1 8 ENTR EXIT
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
DIAG SIGNAL I / O
PREV NEXT JUMP ENTR EXIT
DIAG I / O Test Signals Displayed Here
PREV NEXT JUMP PRNT EXIT
Use the JUMP key to
go directly to a
specific signal
See Appendix A-4 for
a complete list of
available SIGNALS
DIAG I / O JUMP TO: 12
1 2 ENTR EXIT
EXAMPLE:
Enter 12 to Jump to
12) ST_CONC_VALID
DIAG I / O ST_CONC_VALID = ON
PREV NEXT JUMP ON PRNT EXIT
Exit to return
to the
DIAG menu
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG ALRM EXIT
Use theNEXT &PREV
keys to move between
signal types.
Pressing the PRNT key will send a formatted printout to the serial port and can be
captured with a computer or other output device.
EXIT returns
to the main
SAMPLE display

6.13.3. Analog Output Step Test

This test can be used to check the accuracy and proper operation of the analog outputs. The test forces all four

analog output channels to produce signals ranging from 0% to 100% of the full scale range in 20% increments.

This test is useful to verify the operation of the data logging/recording devices attached to the analyzer.

To begin the Analog Output Step Test press:

SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG ALRM EXIT
SAMPLE RANGE = 500.000 PPM CO2 =X.XXX
< TST TST > CAL SETUP
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
DIAG SIGNAL I / O
NEXT ENTR EXIT
Exit-Exit
returns to the
DIAG menu
DIAG
A
NALOG OUTPUT
PREV NEXT ENTR EXIT
DIAG AOUT ANALOG OUTPUT
0% EXIT
Performs
analog output
step test.
0% - 100%
DIAG AOUT ANALOG OUTPUT
[0%] EXIT
Pressing the key under “0%” while performing the test will
pause the test at that level. Brackets will appear around
the value: example: [20%] Pressing the same key again
will resume the test.

SETUP X.X ENTER DIAG PASS: 818

8 1 8 ENTR EXIT