Getting Started Model 360E Instruction Manual
3.1.2. Electrical Connections

3.1.2.1. Power Connection

Attach the power cord to the analyzer and plug it into a power outlet capable of carrying at least 10 A current at
your AC voltage and that it is equipped with a functioning earth ground.
CAUTION
Check the voltage and frequency label on the rear panel of the instrument
(See Figure 3-2 for compatibility with the local power before plugging the
M360E into line power.
Do not plug in the power cord if the voltage or
frequency is incorrect.
CAUTION
Power connection must have functioning ground connection.
Do not defeat the ground wire on power plug.
Turn off analyzer power before disconnecting or
connecting electrical subassemblies.
Do not operate with cover off.
The M360E analyzer can be configured for both 100-130 V and 210-240 V at either 50 or 60 Hz. To avoid
damage to your analyzer, make sure that the AC power voltage matches the voltage indicated on the rear panel
serial number label and that the frequency is between 47 and 63 Hz.
12 05232 Rev B3