Model 360E Instruction Manual Operating Instructions
6.11.2. COMM Port Default Settings
Received from the factory, the analyzer is set up to emulate a DCE or modem, with pin 3 of the DB-9 connector
designated for receiving data and pin 2 designated for sending data.
COM1: RS-232 (fixed) DB-9 male connector.
o Baud rate: 19200 bits per second (baud).
o Data Bits: 8 data bits with 1 stop bit.
o Parity: None.
COM2: RS-232 (configurable), DB-9 female connector.
o Baud rate: 115000 bits per second (baud).
o Data Bits: 8 data bits with 1 stop bit.
o Parity: None.
NOTE
Cables that appear to be compatible because of matching connectors may incorporate
internal wiring that make the link inoperable. Check cables acquired from sources other
than Teledyne Instruments for pin assignments before using.
In its default configuration, the M360E analyzer has two available RS-232 Com ports accessible via 2 DB-9
connectors on the back panel of the instrument. The COM1 connector is a male DB-9 connector and the COM2
is a female DB9 connector.
Male DB-9 (RS-232)
(As seen from outside analyzer)
(DTE mode)
(DCE mode)
1 2 3 4 5
6 7 8 9
RXD GND
TXD
CTS
RTS
1 2 3 4 5
6 7 8 9
TXD GND
RXD
RTS
CTS
Female DB-9 (COM2)
(As seen from outside analyzer)
(DTE mode)
12345
6789
RXD GND
TXD
CTS
RTS
Figure 6-8: Back Panel connector Pin-Outs for COM1 & COM2 in RS-232 mode.
05232 Rev B3 87