Model 360E Instruction Manual APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4
SETUP VARIABLE NUMERIC
UNITS DEFAULT
VALUE VALUE RANGE DESCRIPTION
O2_SLOPE 5 — 1 0.5–2 O2 slope.
O2_OFFSET 5 % 0 -10–10 O2 offset.
RANGE_MODE — SNGL
1 SNGL, DUAL,
AUTO Range control mode.
CONC_RANGE1 Conc. 50 0.1–50000 D/A concentration range 1.
O2_RANGE 5 % 100 0.1–500 O2 concentration range.
RS232_MODE BitFlag 0 0–65535 RS-232 COM1 mode flags. Add
values to combine flags.
1 = quiet mode
2 = computer mode
4 = enable security
8 = enable hardware handshaking
16 = enable Hessen protocol 4
32 = enable multi-drop
64 = enable modem
128 = ignore RS-232 line errors
256 = disable XON / XOFF support
512 = disable hardware FIFOs
1024 = enable RS-485 mode
2048 = even parity, 7 data bits, 1
stop bit
4096 = enable command prompt
BAUD_RATE — 19200
1
300, 1200, 2400,
4800, 9600,
19200, 38400,
57600, 115200
RS-232 COM1 baud rate.
MODEM_INIT “AT Y0 &D0 &H0
&I0 S0=2 &B0 &N6
&M0 E0 Q1 &W0” 1
Any character in the
allowed character set.
Up to 100 characters
long.
RS-232 COM1 modem initialization
string. Sent verbatim plus carriage
return to modem on power up or
manually.
RS232_MODE2 BitFlag 0 0–65535
RS-232 COM2 mode flags.
(Same settings as RS232_MODE.)
BAUD_RATE2 — 19200
1
300, 1200,
2400, 4800,
9600, 19200,
38400, 57600,
115200
RS-232 COM2 baud rate.
MODEM_INIT2 “AT Y0 &D0 &H0
&I0 S0=2 &B0 &N6
&M0 E0 Q1 &W0” 1
Any character in the
allowed character set.
Up to 100 characters
long.
RS-232 COM2 modem initialization
string. Sent verbatim plus carriage
return to modem on power up or
manually.
RS232_PASS Password 940331 0–999999 RS-232 log on password.
MACHINE_ID ID 320 0–9999 Unique ID number for instrument.
COMMAND_PROMPT — “Cmd>
1
Any character in the
allowed character set.
Up to 100 characters
long.
RS-232 interface command prompt.
Displayed only if enabled with
RS232_MODE variable.
TEST_CHAN_ID — NONE
1
NONE,
CO MEASURE,
CO REFERENCE,
VACUUM PRESSURE,
Diagnostic analog output ID.
05233 Rev G.4 A-19