Operating Instructions Model 360E Instruction Manual
RANGE
RANGE11
RANGE21
02 RANGE
STABIL
MEAS
REF
MR RATIO
PRES
SAMP FL
SAMP TEMP
BENCH TEMP
WHEEL TEMP
BOX TEMP
PHT DRIVE
SLOPE
OFFSET
TEST
TIME
SAMPLE RANGE = 500.000 PPM CO2 = XXX.X
< TST TST > CAL SETUP
1
Only appears instrument is set
for DUAL or AUTO reporting
range modes.
2 Only appears if 02 Sensor
Option is installed.
Refer to
Table 6-2 for
definitions of
these test
functions.
Toggle <TST TST> keys to
scroll throu
g
h list of functions
Figure 6-3: Viewing M360E TEST Functions
NOTE
A value of “XXXX” displayed for any of the TEST functions indicates an out-of-range
reading or the analyzer’s inability to calculate it.
All pressure measurements are represented in terms of absolute pressure. Absolute,
atmospheric pressure is 29.92 in-Hg-A at sea level. It decreases about 1 in-Hg per 300
m gain in altitude. A variety of factors such as air conditioning and passing storms can
cause changes in the absolute atmospheric pressure.
54 05232 Rev B3