Model 360E Instruction Manual APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4
SETUP VARIABLE NUMERIC
UNITS DEFAULT
VALUE VALUE RANGE DESCRIPTION
CO_DWELL Seconds 0.2 0.1–30
Dwell time before taking measure or
reference sample.
CO_SAMPLE Samples 1 1–30
Number of samples to take in
measure or reference mode.
FILT_SIZE Samples 750 1–1000 Moving average filter size.
FILT_ASIZE Samples 48 1–1000
Moving average filter size in adaptive
mode.
FILT_DELTA PPM 4 1–1000
Absolute change to trigger adaptive
filter.
FILT_PCT % 10 1–100
Percent change to trigger adaptive
filter.
FILT_DELAY Seconds 90 0–180
Delay before leaving adaptive filter
mode.
FILT_ADAPT — ON ON, OFF
ON enables adaptive filter; OFF
disables it.
O2_DWELL 5 Seconds 1 0.1–30
Dwell time before taking each
sample.
O2_FILT_ADAPT 5 — ON ON, OFF
ON enables O2 adaptive filter; OFF
disables it.
O2_FILT_SIZE 5 Samples 60 1–500
O2 moving average filter size in
normal mode.
O2_FILT_ASIZE 5 Samples 10 1–500
O2 moving average filter size in
adaptive mode.
O2_FILT_DELTA 5 % 2 0.1–100
Absolute change in O2 concentration
to shorten filter.
O2_FILT_PCT 5 % 2 0.1–100
Relative change in O2 concentration
to shorten filter.
O2_FILT_DELAY 5 Seconds 20 0–300
Delay before leaving O2 adaptive
filter mode.
O2_DIL_FACTOR 5 — 1 0.1–1000
Dilution factor for O2. Used only if is
dilution enabled with
FACTORY_OPT variable.
USER_UNITS — PPM
1 PPB,PPM,
UGM,MGM
Concentration units for user
interface.
DIL_FACTOR — 1 0.1–1000
Dilution factor. Used only if is dilution
enabled with FACTORY_OPT
variable.
DARK_CAL_DURATION Seconds 60 10–600
Duration of dark cal. First two-thirds
is stabilization period; final third is
measure period.
DARK_MEAS_MV mV 0 -1000–1000 Dark offset for measure reading.
DARK_REF_MV mV 0 -1000–1000 Dark offset for reference reading.
LIN_TARGET_CONC1 Conc 300 1–10000
Target concentration during linearity
adjustment for range 1.
LIN_NORM_CONC1 PPM 300 0.01–10000
Target concentration during linearity
adjustment normalized for T/P for
range 1.
LIN_RATIO1 — 1 0.01–100
Measure/reference ratio measured
during linearity adjustment for range
1.
LIN_CORRECT1 1 0.001–999.999 Linearity correction factor for range 1.
LIN_TARGET_CONC2 Conc 300 1–10000
Target concentration during linearity
adjustment for range 2.
LIN_NORM_CONC2 PPM 300 0.01–10000
Target concentration during linearity
adjustment normalized for T/P for
range 2.
05233 Rev G.4 A-17