Model 360E Instruction Manual Operating Instructions
NOTE
Upper span limit setting for the individual range modes are shared. Resetting the span
limit in one mode also resets the span limit for the corresponding range in the other
modes as follows:
SNGL DUAL AUTO
Range  Range1  Low Range
Range2  High Range
6.8.3. Single Range mode (SNGL)
This is the default reporting range mode for the analyzer. In single range mode both A1 and A2 are set to the
same reporting range. This reporting range can be any value between 50 ppb and 2 000 ppm.
While the two outputs always have the same reporting range, the span, signal offset and scaling of their
electronic signals may be configured for differently (e.g., A1 = 0-10 V; A2 = 0-0.1 V). See Section 6.13.4 for
instructions on adjusting these parameters.
To select SNGL range mode and to set the upper limit of the range, press:
SETUP C.3 PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SETUP C.3 RANGE CONTROL MENU
MODE
SET UNIT EXIT
SETUP C.3 RANGE MODE: SNGL
SNGL
DUAL AUTO ENTR EXIT
EXIT x 2 returns
to the main
SAMPLE display
SETUP C.3 RANGE MODE: SNGL
SNGL
DUAL AUTO ENTR EXIT
SETUP C.3 RANGE CONTROL MENU
MODE
SET UNIT EXIT
SETUP C.3 RANGE: 500.0 Conc
0 0 5 0 0 .0 ENTR EXIT
SETUP C.3 RANGE CONTROL MENU
MODE
SET UNIT EXIT
SAMPLE* RANGE = 500.000 PPM CO2 X.XXX
< TST TST > CAL SETUP
05232 Rev B3 79