Model 360E Instruction Manual Operating Instructions
6.2.2. Warning Messages
The most common instrument failures will be reported as a warning on the analyzer’s front panel and through
the COM ports. Section 11.1.1 explains how to use these messages to troubleshoot problems. Section 3.2.3
shows how to view and clear warning messages.
Table 6-3 lists all warning messages for the current version of software.
Table 6-3: List of Warning Messages
MESSAGE MEANING
ANALOG CAL WARNING The instrument’s A/D circuitry or one of its analog outputs is not calibrated.
BENCH TEMP WARNING The Temperature of the optical bench is outside the specified limits.
BOX TEMP WARNING Remote span calibration failed while the dynamic span feature was set to turned on
CANNOT DYN SPAN Remote zero calibration failed while the dynamic zero feature was set to turned on
CANNOT DYN ZERO Configuration was reset to factory defaults or was erased.
CONC ALRM1 WARNING Concentration alarm 1 is enabled and the measured CO2 level is the set point.
CONC ALRM2 WARNING Concentration alarm 2 is enabled and the measured CO2 level is the set point.
CONFIG INITIALIZED Configuration storage was reset to factory configuration or erased.
DATA INITIALIZED iDAS data storage was erased.
O2 CELL TEMP WARN O2 sensor cell temperature outside of warning limits.
PHOTO TEMP WARNING The temperature of the IR photometer is outside the specified limits.
REAR BOARD NOT DET The CPU is unable to communicate with the motherboard.
RELAY BOARD WARN The firmware is unable to communicate with the relay board.
SAMPLE FLOW WARN The flow rate of the sample gas is outside the specified limits.
SAMPLE PRESS WARN Sample gas pressure outside of operational parameters.
SAMPLE TEMP WARN The temperature of the sample gas is outside the specified limits.
SOURCE WARNING The IR source may be faulty.
SYSTEM RESET The computer was rebooted.
WHEEL TEMP WARNING The Gas Filter Correlation wheel temperature is outside the specified limits.