Model 360E Instruction Manual Getting Started
3. GETTING STARTED

3.1. Unpacking and Initial Set Up

CAUTION
To avoid personal injury, always use two persons to lift and carry the Model 360E.
1. Verify that there is no apparent external shipping damage. If damage has occurred, please advise the
shipper first, then Teledyne Instruments.
2. Included with your analyzer is a printed record of the final performance characterization performed on
your instrument at the factory. This record, titled Final Test and Validation Data Sheet (P/N 04596) is an
important quality assurance and calibration record for this instrument. It should be placed in the quality
records file for this instrument.
3. Carefully remove the top cover of the analyzer and check for internal shipping damage.
Remove the set-screw located in the top, center of the Front panel.
Remove the 2 screws fastening the top cover to the unit (one per side towards the rear).
Slide the cover backwards until it clears the analyzer’s front bezel.
Lift the cover straight up.
NOTE
Some versions of the 360E CO2 Analyzer may have a spring loaded fastener at the top
center of the rear panel and as many as eight screws (four per side) fastening the top
cover to the chassis.
NOTE
Printed circuit assemblies (PCAs) are sensitive to electro-static discharges too small to
be felt by the human nervous system. Failure to use ESD protection when working with
electronic assemblies will void the instrument warranty.
See Chapter 12 for more information on preventing ESD damage.
CAUTION
Never disconnect PCAs, wiring harnesses or electronic subassemblies while under
power.
4. Inspect the interior of the instrument to make sure all circuit boards and other components are in good
shape and properly seated.
05232 Rev B3 7