Model 360E Instruction Manual APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4
SETUP VARIABLE NUMERIC
UNITS DEFAULT
VALUE VALUE RANGE DESCRIPTION
O2_FILT_DELAY 5 Seconds 20 0–300
Delay before leaving O2 adaptive
filter mode.
O2_DIL_FACTOR 5 — 1 0.1–1000
Dilution factor for O2. Used only if is
dilution enabled with
FACTORY_OPT variable.
USER_UNITS — PPM
1 PPB,PPM,
UGM,MGM
Concentration units for user
interface.
DIL_FACTOR — 1 0.1–1000
Dilution factor. Used only if is dilution
enabled with FACTORY_OPT
variable.
DARK_CAL_DURATION Seconds 60 10–600
Duration of dark cal. First two-thirds
is stabilization period; final third is
measure period.
DARK_MEAS_MV mV 0 -1000–1000 Dark offset for measure reading.
DARK_REF_MV mV 0 -1000–1000 Dark offset for reference reading.
LIN_TARGET_CONC1 Conc 300 1–10000
Target concentration during linearity
adjustment for range 1.
LIN_NORM_CONC1 PPM 300 0.01–10000
Target concentration during linearity
adjustment normalized for T/P for
range 1.
LIN_RATIO1 — 1 0.01–100
Measure/reference ratio measured
during linearity adjustment for range
1.
LIN_CORRECT1 1 0.001–999.999 Linearity correction factor for range 1.
LIN_TARGET_CONC2 Conc 300 1–10000
Target concentration during linearity
adjustment for range 2.
LIN_NORM_CONC2 PPM 300 0.01–10000
Target concentration during linearity
adjustment normalized for T/P for
range 2.
LIN_RATIO2 — 1 0.01–100
Measure/reference ratio measured
during linearity adjustment for range
2.
LIN_CORRECT2 1 0.001–999.999 Linearity correction factor for range 2.
LIN_TARGET_CONC Conc 300 1–10000
Target concentration during linearity
adjustment.
LIN_NORM_CONC PPM 300 0.01–10000
Target concentration during linearity
adjustment normalized for T/P.
LIN_RATIO — 1 0.01–100
Measure/reference ratio measured
during linearity adjustment.
LIN_CORRECT 1 0.001–999.999 Linearity correction factor.
CO2_COMP_ENABLE — OFF ON, OFF
ON enables CO2 compensation; OFF
disables it.
CO2_COMP_CONC % 0 0–20
CO2 concentration to compensate
for.
CO_CONST1 500 100–50000 CO calculation constant.
CO_CONST2 1.448 0–10 CO calculation constant.
ET_MEAS_GAIN — 1 0.0001–9.9999
Electrical test gain factor for measure
reading.
ET_REF_GAIN — 1 0.0001–9.9999
Electrical test gain factor for
reference reading.
ET_TARGET_DET mV 4375 0–5000
Target detector reading during
electrical test.
ET_TARGET_CONC PPM 40 1–10000
Target concentration during electrical
test.
ET_CONC_RANGE Conc. 50 0.1–50000
D/A concentration range during
electrical test.
05233 Rev G.4 A-11