Operating Instructions Model 360E Instruction Manual

6.13.4.5. AIN Calibration

This is the sub-menu to conduct the analog input calibration. This calibration should only be necessary after

major repair such as a replacement of CPU, motherboard or power supplies. Activate the ANALOG I/O

CONFIGURATION MENU (See Section 6.13.1), then press:

Exit at any time to
return to the main
DIAG menu
DIAG
A
NALOG I / O CONFIGURATION
PREV NEXT ENTR EXIT
Continue pressing SET? until …
DIAG AIO
A
IN CALIBRATED: NO
< SET SET> CAL EXIT
DIAG AIO AIN CALIBRATED:YES
< SET SET> CAL EXIT
Instrument
calibrates
automatically
Exit to return to the
ANALOG I/O
CONFIGURATION
MENU
STARTING FROM ANALOG I / O CONFIGURATION MENU
DIAG AIO CALIBRATING A/D ZERO
CALIBRATING A/D SPAN

116 05232 Rev B3