4.2 Instrument Status

The instrument’s current status condition is displayed in decimal format on the front panel HOME screen and in the Status column of the data download. The status condition relates to various subcomponents of the instrument, e.g., Detector, Flow, LED, etc. The decimal number represents a sum of all of the status conditions occurring at any given time. (Multiple status conditions are interpreted using subtraction of the largest possible Status Condition value using the table below.

Table 1. Status Conditions and Descriptions.

 

Status

 

Parameter

Condition

Description

Detector

0

measuring

 

1

not measuring (due to either tape advance, fast

 

 

calibration in progress)

 

 

 

 

2

calibrating (LED, Flow, Tape Sensors)

 

 

 

 

3

stopped

Flow

 

flow OK

 

 

 

 

4

flow low/high by more than 0.25 LPM

 

 

 

 

8

Flow check status history

 

12

Flow L/H and check status history

LED

 

LEDs OK

 

16

calibrating

 

32

calibration error in one or more channels

 

 

(at least one channel OK)

 

48

led error (all channels calibration error, COM error)

 

 

 

Tape

 

tape advance OK

Advance

128

tape warning (less than 10 spots left)

 

256

tape last warning (card box visible, less than 5 spots left)

 

 

 

 

384

tape error (tape not moving, end of tape)

 

 

 

Tests

 

No test

 

1024

Stability test

 

2048

Clean air test

 

 

 

 

3072

Change tape

 

 

 

When Status Condition 3 is encountered, the Aethalometer stops. In all other statuses, it continues to operate with a warning, and the data is flagged accordingly. The status is rep‐ resented by one value, which can point to one parameter or a combination of parameters.

07645A DCN6170