5.1 Cleaning the cyclone

Model 633 Aethalometer® Black Carbon Monitor User Manual

Teledyne Advanced Pollution Instrumentation

5.1 Cleaning the cyclone

Remove the cyclone and disassemble (shown in the following instructions), cleaning the various parts thoroughly. The recommended cleaning method is immersion in an ultrasonic cleaner with water and mild soap. However, wiping with a water‐dampened and lint‐free cloth is sufficient.

IMPORTANT: Protracted soaking in soap/caustic solutions will damage the aluminum components.

Do not re­aerosolize hazardous materials when using compressed air for cleaning.

Cyclone cleaning procedure:

1. Remove cyclone assembly from instrument.

2. Remove inlet cover.

48