Teledyne Advanced Pollution Instrumentation

Model 633 Aethalometer® Black Carbon Monitor User Manual

3.Disconnect flowmeter from the input.

4.Connect the flow meter to the output of the instrument by a long tube (10 m) to reduce the oscillations, which can influence the flow measurements.

5.Take a Fout reading.

IMPORTANT!

If you use a volumetric flow meter take into account the difference in air pressure and temperature between the flow in and flow out (use the ideal gas law equation).

Report

 

 

 

 

 

 

 

 

 

 

 

Aethalometer

Flowmeter

Date

Fin (LPM)

Fout (LPM)

ζ (%)

SN

 

 

 

 

 

 

 

 

 

 

 

AE33‐S00‐0054

TSI4100

15.1.2013

4.67

5.01

6.8

 

 

 

 

 

 

5.3 Automatic flow meter calibration

First, ensure that a leak test is completed (Section 5.1) prior to running this calibration procedure. For the automatic calibration a TSI4100 flow meter, data cable and the flow calibration pad are needed.

Before calibration check TSI4100 flow meter settings:

07645A DCN6170

51

Page 53
Image 53
Teledyne 633 user manual Report