Model 633 Aethalometer® Black Carbon Monitor User Manual

Teledyne Advanced Pollution Instrumentation

7. Remove allen cap screws from cyclone top.

8.Clean all parts and thoroughly dry; then reassemble in reverse order.

5.2 Leak test

Always conduct a leak test prior to running a flow meter calibration, whether automatic (section 5.2) or manual (section 5.3).

Leakage (ζ) is measured during instrument operation. It is calculated from the flow in (Fin) and flow out (Fout).

ζ=1‐(Fin/Fout)

Average leakage is 7% at 5 LPM. It can differ slightly from spot to spot and during the spot loading.

Flow measurement using a mass‐flow meter (for example TSI4100)

1.Connect the flow meter to the input of the instrument.

2.Wait for a few seconds for the flow to stabilize and read the flow Fin.

50

07645A DCN6170

Page 52
Image 52
Teledyne 633 user manual Leak test