Model 633 Aethalometer® Black Carbon Monitor User Manual

Tape sensor calibration procedure:

3.The Tape Sensors Adjustment screen prompts to place the small tapeholder [cylinder] on left and large on right, then press OK. Secure with tape retention disks (not shown in illustration) before pressing OK!

4.The next prompt is to switch the tapeholders [cylinder], then press OK. Secure with tape retention disks (not shown in illustration) before pressing OK!

5.Last screen shows that the adjustment is complete.

6.After you’ve pressed either OK or Cancel, return the sampling tape per the applicable steps in Section 3.4 Filter tape installation).