5 MAINTENANCE and SERVICE

The Aethalometer needs to be maintained regularly. We recommend the following schedule for regular maintainance.

Item

Sample inlet flow

Inspect the sample line tubing

Inspect and clean the size selective inlet (if present)

Clean cyclone

(5.1 Cleaning the cyclone)

Verify time and date (if not set to update automatically) (Time & Date in OPERATION>GENERAL page).

Inspect optical chamber, clean if necessary

(Section 5.6 Cleaning the optical head)

Check flow, calibrate if necessary

(Section 5.2 Leak test).

(For calibration: Section 5.3 Automatic flow meter calibration or Section 5.4 Manual flow meter calibration).

Change by‐pass cartridge filter

(Section 5.7 Changing the bypass cartridge filter)

Install a new filter tape roll

(Section 3.4 Filter tape installation)

Calibrate tape sensor

(Section 5.8 Calibrating the tape sensor)

Frequency

Monthly

Monthly

Monthly

Weekly at first to gauge debris accumulation, then adjust as needed.

Monthly

Semi‐annually

Site dependent, use educated judgement!

Semi‐annually or as needed.

Annually or as needed.

As needed. The instrument issues a warning.

When measurement chamber has been removed/replaced.

07645A DCN6170

47

Page 49
Image 49
Teledyne 633 user manual Maintenance and Service, Frequency