Manual flow calibration procedure:

10.In the display’s blank FLOW field enter the flow value from your flow‐meter measured in or converted to milliliters (this will be approximately 3000; do NOT change the pressure or temperature settings), and select OK.

11.Wait for the pump to run for several seconds to achieve a stable flow.

12.In the display’s blank FLOW field enter the flow value from your flow‐meter measured in or converted to milliliters (this will be approximately 5000; do NOT change the pressure or temperature settings), and select OK.