5.4 Manual flow meter calibration

Teledyne Advanced Pollution Instrumentation

Model 633 Aethalometer® Black Carbon Monitor User Manual

5.4 Manual flow meter calibration

First, ensure that a leak test is completed (Section 5.1) prior to running this calibration procedure. For the manual calibration a flow meter, data cable and the flow calibration pad are needed. Please note flow measurements are in Milliliters, and maximum flow rate should be no greater than 5 LPM (5,000 milliliters).

Manual flow calibration procedure:

1.STOP measurements in the

OPERATION>GENERAL

menu.

2.In the

OPERATION>ADVANCED menu press FLOWCAL.

3.Select MANUAL to start the manual flow calibration.