Oxygen Analyzer

Installation

 

 

 

Zero:

Floating input—5 to 24 V input across the + and – pins puts the analyzer into the Zero mode. Either side may be grounded at the source of the signal. 0 to 1 volt across the terminals allows Zero mode to terminate when done. A synchronous signal must open and close the external zero valve appropriately. See Remote Probe Connector. (The –C option internal valves operate automatically).

Span:

Floating input—5 to 24 V input across the + and – pins puts the analyzer into the Span mode. Either side may be grounded at the source of the signal. 0 to 1 volt across the terminals allows Span mode to terminate when done. A synchronous signal must open and close external span valve appropriately. See Figure 3-5 Remote Probe Connector. (The –C option internal valves operate automatically.)

Cal Contact:

This relay contact is closed while analyzer is spanning and/or zeroing. (See Remote Calibration Protocol below.)

Table 3-3: Remote Calibration Connections

Pin Function

9+ Remote Zero

11 – Remote Zero

10 + Remote Span

12 – Remote Span

40 Cal Contact

41 Cal Contact

Remote Calibration Protocol: To properly time the Digital Remote Cal Inputs to the Model BDS 3960 Analyzer, the customer's controller must monitor the Cal Relay Contact.

Teledyne Analytical Instruments

37

Page 37
Image 37
Teledyne BDS-3960 operating instructions Zero, Cal Contact