Maintenance

BDS 3960

 

 

 

 

 

 

System. Pressing this button will open a second screen that consists of several variables that regulate the internal operations of the analyzer.

Training Video. This buttons launches a video player. At this point you could view a short training video regarding BDS technology. The presentation will time out and return back to the Analyzer application after a period of time.

Ranges. The list box is used to set the analyzer on a specific range as well as set the limits of the ranges.

Any function can be selected at any time by pressing the

appropriate button. The order as presented in this manual is appropriate for an initial setup.

Each of these functions is described in greater detail in the following procedures.

4.3The System Screen

Pressing the System button on the main screen will bring the System screen.

The System screen will time out in five minutes and return to the main screen. Pressing the Done button will also return you to the main screen.

4.3.1Communication Information and Calibration Parameters

The first block on the System screen displays information about data received, calibration parameters, and data sent to the I/O hardware.

Received data: 014508900

Zero Offset: 0

Span Factor: 38.00

Amp Gain/Range ID/DAC: 0/0/265

The first line displays the data received from the I/O PCB. It consists of nine digits. The first four digits are the Analog to Digital counts of the BDS sensor amplifier. The range of this count is between 0000 and 4095. The following four digits are the Analog to Digital counts of the temperature

Teledyne Analytical Instruments

44

Page 44
Image 44
Teledyne BDS-3960 operating instructions System Screen, Communication Information and Calibration Parameters