Installation

BDS 3960

 

 

 

 

 

 

Installation

Installation of the Model BDS 3960 Analyzer includes:

Unpacking

Mounting

Gas connections

Electrical connections

Filling the Sensor with Electrolyte.

Testing the system.

3.1Unpacking the Analyzer

Although the analyzer is shipped complete, certain parts, such as the electrolyte, are wrapped separately to be installed on site as part of the installation. Carefully unpack the analyzer and inspect it for damage. Immediately report any damage or shortages to the shipping agent.

3.2Mounting the Analyzer

The Model BDS 3960 is for indoor use in a general purpose area. It is NOT for hazardous environments of any type.

The standard model is designed for flush panel mounting. Figure 3-1 is an illustration of the BDS 3960 standard front panel and mounting bezel. There are four mounting holes—one in each corner of the rigid frame. The drawings section in the rear of this manual contains outline dimensions and mounting hole spacing diagrams.

Teledyne Analytical Instruments

30

Page 30
Image 30