Installation

BDS 3960

 

 

 

 

 

 

7.Pour the rest of the electrolyte into the sensor. Gas bubbles in the top portion of the sensor do not affect the sensor performance. One bottle of electrolyte is sufficient to rise the electrolyte level to the MAX line. For the rest of sensor life, no further electrolyte addition is needed.

8.Install the sensor cap.

9.Do not connect the sensor's electric connector at this stage.

Figure 3.5: Adding Electrolyte to the BDS Sensor

3.5Testing the System

Before plugging the instrument into the power source:

Check the integrity and accuracy of the gas connections. Make sure there are no leaks. Gas connection should allow for some movement on the drawer assembly so that the pressure regulator and the sensor connector cable can be reached inside.

Check the integrity and accuracy of the electrical connections. Make sure there are no exposed conductors

Check pressure of sample gas as well as instrument air. Set the internal pressure regulator fully counterclockwise.

Teledyne Analytical Instruments

40

Page 40
Image 40
Teledyne BDS-3960 operating instructions Testing the System, Adding Electrolyte to the BDS Sensor