CHAPTER 7 RS232 & RS485 REMOTE CONTROL

7.1 INTRODUCTION

This Chapter describes the operation of the GenesysTM 3300W power supplies via the serial communication port. Details of the initial set-up, operation via RS232 or RS485, the command set and the communication protocol are described in this Chapter.

7.2 CONFIGURATION

7.2.1 Default setting

The power supply is shipped with the following settings:

-Address

6

-Output

Off

-Baud-rate

9600

-Start up mode

Safe-start

-RS232/485

RS232

-OVP

Maximum

-Vout setting

0

-UVL

0

-Iout setting

Maximum

-Foldback

Off

-Master/Slave

H1 (Master)

-Front panel:

Unlocked (UFP)

7.2.2Address setting

The power supply address can be set to any address between 0 and 30. Follow the instructions described below to set the unit address.

1.If the unit is in Remote mode (front panel REM/LOC LED illuminated), press the REM/LOC button to put the unit into Local mode.

2.Press and hold for the REM/LOC button for approximately 3 sec. The VOLTAGE display will indicate the unit address.

3.Using the VOLTAGE adjust encoder, select the unit address.

To preview the power supply address at any time, press and hold the REM/LOC button for approx. 3 sec. The VOLTAGE display will indicate the power supply address.

7.2.3RS232 or RS485 selection

To select between RS232 or RS485 set the rear panel setup switch SW1-6 position to: -DOWN for RS232

-UP for RS485

7.2.4Baud Rate setting

Five optional Baud rates are possible: 1200, 2400, 4800, 9600 and 19200. To select the desired rate, the following steps should be taken:

1.If the unit is in Remote mode (front panel REM/LOC LED illuminates), press REM/LOC button to put the unit into Local mode.

2.Press and hold the REM/LOC button for approximately 3 sec. The CURRENT display will show the unit Baud Rate.

3.Using the CURRENT adjust encoder, select the desired Baud Rate.

7.2.5Setting the unit into Remote or Local mode

1. The unit will be put into Remote mode only via serial communication command. Commands that will put the unit into Remote mode are:

RST

PV n

OUT n

PC n

RMT n

 

(for n values see Tables 7-5 and 7-7)

50

83-507-5002 Rev. B

Image 58
Contents Genesystm 750W Half Rack Page Page Genesys Manual Supplement Table of Contents Front and Rear Panel Controls and CONNECTORS………... ……… ……… Isolated Analog Progamming Option ……………………………..…… Warranty Safety Instructions Safety Instructions SICHERHEITS-INSTALLATIONS Anweisungen Anderungen and Bauteileersatz Sicherheits-und WarnsymboleUmweltbedingungen Introduction Features and optionsGeneral Information General DescriptionAnalog voltage programming and monitoring Multiple output power systemControl via the serial communication port Parallel operationAccessories Specifications ChapterFront Panel Supplemental Characteristics Genesystm Genh 750W Power Supplies Outline Drawings Preparation for USE InstallationGeneral Initial InspectionsSingle unit installation Rack MountingAC Source Requirements AC Input Power ConnectionLOCATION, Mounting and Cooling GENH/RM Rack Mounting outline drawingsAC Input Cord TURN-ON Checkout ProcedureAC Input Connector GeneralOVP Check Constant Voltage CheckConstant Current Check UVL CheckConnecting the Load Wire termination AWGMaking the load connections Noise and Impedance EffectsInductive loads 6V to 60V ModelsBus-bars shield mounting 80V to 600V Models 83-507-5002 Rev. B Connecting single loads, remote sensing Connecting single loads, local sensing defaultConnecting multiple loads, radial distribution method Multiple load connection with distribution terminals Grounding outputsLocal and Remote Sensing Sense wiringOutput Terminal Grounding Local sensing10.4 J2 sense connector technical information Remote sensingFront Panel Controls and Indicators Front and Rear Panel Controls and ConnectorsVoltage Auxiliary Function Rear Panel Connections and Controls Rear panel connections and controls Number DescriptionRear Panel SW1 Setup Switch 1 SW1 position functionResetting the SW1 switch Rear Panel J1 Programming and Monitoring ConnectorMaking J1 connections 83-507-5002 Rev. B Ifcom SignalEnain LocalLocal Operation Constant Voltage ModeConstant Current Mode Standard OperationOver Voltage Protection OVP Setting the OVP levelResetting the OVP circuit Automatic CrossoverFoldback Protection Output SHUT-OFF SO Control VIA Rear Panel J1 ConnectorUnder Voltage Limit UVL Output ON/OFF ControlENABLE/DISABLE Control VIA Rear Panel J1 Connector CV/CC SignalLast Setting Memory Psok SignalOver Temperature Protecton OTP SAFE-START and AUTO-RESTART ModesSeries Operation Series connection for increased output voltageSeries connection for positive and negative output voltage Series connection for positive/negative output voltagesParallel Operation Setting up the Master unitSetting up the Slave units Basic parallel operationAdvanced parallel configuration Setting Over Voltage protectionSetting Foldback protection Setting the units as Master or SlaveCurrent display accuracy Master and Slave units default operationTo release units from Slave mode AST OFF OUT OFFDAISY-CHAIN Connection Front Panel LockingUnlocked front panel Locked front panelRemote Analog Programming LOCAL/REMOTE Analog IndicationDown Remote voltage programming connectionResisitive Programming of Output Voltage and Current Limit Remote resistive programmingRemote Monitoring of Output Voltage and Current RS232 & RS485 Remote Control Configuration6 RS232/RS485 port in Local mode Rear Panel RS232/RS485 ConnectorFront panel in Remote mode MD Mode SRQ MD Mode DescriptionMD Mode enable Serial communication mode MD Mode SRQ RetransmissionConnecting Power Supplies to RS232 or RS485 BUS Single power supplyCommunication Interface Protocol Error Messages Command SET DescriptionID Control Commands Initialization Control CommandsOutput Control Commands FILTER? DVC?Filter OUT?Global Output Commands Disable SRQ retransmission MD Mode Option Required Single Byte CommandsEnable MD Mode MD Mode Option Required Enable SRQ retransmission MD Mode Option RequiredPrint Power On Time Addressed commands with response Read registersRetransmit Last Message Addressed commands without response Acknowledge SRQ Test if MD Mode is InstalledRe-enable SRQ with out reading/clearing the Seve Register 110x STATUS, Error and SRQ Registers Status Control CommandsConditional Registers TXDService Request Enable and Event Registers BITFault Enable Register Fault Event RegisterStatus Enable Register Status Event RegisterSerial Communication Test SET-UP Power supply set-upCommunication Test 1 0-5V/0-10V Option PN IS510 SpecificationsIsolated Analog Programming Option PPM/CIsolated Programming & Monitoring Connector Setup and Operating Instructions Maintenance OUP Fuse Rating