CHAPTER 6 REMOTE ANALOG PROGRAMMING

6.1 INTRODUCTION

The rear panel connector J1 allows the user to program the power supply Output Voltage and Current with an analog device. J1 also provides monitoring signals for Output Voltage and Out- put Current. The programming range and monitoring signals range can be selected between 0- 5V or 0-10V using the setup switch SW1. When the power supply is in Remote Analog pro- gramming mode, the serial communication port is active and can be used to query the power supply settings.

CAUTION

COM (J1-12), VPGM_RTN (J1-22) AND IPGM_RTN (J1-23) terminals of J1 connect internally to the -Sense potential (-S). Do not connect these terminals to any potential other than -Sense (-S), as it may damage the power supply.

6.2 LOCAL/REMOTE ANALOG INDICATION

Contact 8 of J1 (Fig.4-2, Item 5) accepts TTL signal or Open-Short contact (referenced to J1-12) to select between Local or Remote Analog programming of the Output Voltage and Current.

In Local mode, the Output Voltage and Output Current can be programmed via the front panel VOLTAGE and CURRENT encoders or via the RS232/RS485 port. In Remote Analog mode, the Output Voltage and current can be programmed by analog voltage or by programming resistors via J1 contacts 9 and 10 (refer to Sections 6.4 and 6.5). Refer to Table 6-1 for Local/Remote Analog control (J1-8) function and Setup switch SW1-1, 2 setting.

SW1-1, 2 setting

J1-8 function

Output Voltage/

Current setting

 

 

Down (default)

No effect

Local

Up

“0” or Short

Remote

“1” or Open

Local

 

Table 6-1: Local/Remote Analog control function

6.3 LOCAL/REMOTE ANALOG INDICATION

Contact 21 of J1 (Fig. 4-2, Item 5) is an open collector output that indicates if the power supply is in Local mode or in Remote Analog mode. To use this output, connect a pull-up resistor to a volt- age source of 30Vdc maximum. Choose the pull-up resistor so that the sink current will be less than 5mA when the output is in a low state. Refer to table 6-2 for J1-21 function.

J1-8

SW1-1

SW1-2

J1-21 signal

Mode

 

 

 

 

 

TTL “0” or short

Down

Down

Open

Local (FP)

 

Down

Up

0i0.6V

Remote Ana-

 

 

 

 

log

 

Up

Down

0i0.6V

Remote Ana-

 

 

 

 

log

 

Up

Up

0i0.6V

Remote Ana-

 

 

 

 

log

TTL “1” or open

Down or Up

Down or Up

Open

Local (FP)

Table 6-2: Local/Remote Analog indication

46

83-507-5002 Rev. B

Image 54
Contents Genesystm 750W Half Rack Page Page Genesys Manual Supplement Table of Contents Front and Rear Panel Controls and CONNECTORS………... ……… ……… Isolated Analog Progamming Option ……………………………..…… Warranty Safety Instructions Safety Instructions SICHERHEITS-INSTALLATIONS Anweisungen Sicherheits-und Warnsymbole Anderungen and BauteileersatzUmweltbedingungen Introduction Features and optionsGeneral Information General DescriptionAnalog voltage programming and monitoring Multiple output power systemControl via the serial communication port Parallel operationAccessories Specifications ChapterFront Panel Supplemental Characteristics Genesystm Genh 750W Power Supplies Outline Drawings Preparation for USE InstallationGeneral Initial InspectionsSingle unit installation Rack MountingAC Source Requirements AC Input Power ConnectionLOCATION, Mounting and Cooling GENH/RM Rack Mounting outline drawingsAC Input Cord TURN-ON Checkout ProcedureAC Input Connector GeneralOVP Check Constant Voltage CheckConstant Current Check UVL CheckConnecting the Load Wire termination AWGMaking the load connections Noise and Impedance EffectsInductive loads 6V to 60V ModelsBus-bars shield mounting 80V to 600V Models 83-507-5002 Rev. B Connecting single loads, local sensing default Connecting single loads, remote sensingConnecting multiple loads, radial distribution method Multiple load connection with distribution terminals Grounding outputsLocal and Remote Sensing Sense wiringOutput Terminal Grounding Local sensing10.4 J2 sense connector technical information Remote sensingFront and Rear Panel Controls and Connectors Front Panel Controls and IndicatorsVoltage Auxiliary Function Rear Panel Connections and Controls Rear panel connections and controls Number DescriptionRear Panel SW1 Setup Switch 1 SW1 position functionRear Panel J1 Programming and Monitoring Connector Resetting the SW1 switchMaking J1 connections 83-507-5002 Rev. B Ifcom SignalEnain LocalLocal Operation Constant Voltage ModeConstant Current Mode Standard OperationOver Voltage Protection OVP Setting the OVP levelResetting the OVP circuit Automatic CrossoverFoldback Protection Output SHUT-OFF SO Control VIA Rear Panel J1 ConnectorUnder Voltage Limit UVL Output ON/OFF ControlENABLE/DISABLE Control VIA Rear Panel J1 Connector CV/CC SignalLast Setting Memory Psok SignalOver Temperature Protecton OTP SAFE-START and AUTO-RESTART ModesSeries Operation Series connection for increased output voltageSeries connection for positive and negative output voltage Series connection for positive/negative output voltagesParallel Operation Setting up the Master unitSetting up the Slave units Basic parallel operationAdvanced parallel configuration Setting Over Voltage protectionSetting Foldback protection Setting the units as Master or SlaveCurrent display accuracy Master and Slave units default operationTo release units from Slave mode AST OFF OUT OFFDAISY-CHAIN Connection Front Panel LockingUnlocked front panel Locked front panelRemote Analog Programming LOCAL/REMOTE Analog IndicationDown Remote voltage programming connectionResisitive Programming of Output Voltage and Current Limit Remote resistive programmingRemote Monitoring of Output Voltage and Current RS232 & RS485 Remote Control ConfigurationRear Panel RS232/RS485 Connector 6 RS232/RS485 port in Local modeFront panel in Remote mode MD Mode SRQ MD Mode DescriptionMD Mode enable Serial communication mode MD Mode SRQ RetransmissionConnecting Power Supplies to RS232 or RS485 BUS Single power supplyCommunication Interface Protocol Error Messages Command SET DescriptionInitialization Control Commands ID Control CommandsOutput Control Commands FILTER? DVC?Filter OUT?Global Output Commands Disable SRQ retransmission MD Mode Option Required Single Byte CommandsEnable MD Mode MD Mode Option Required Enable SRQ retransmission MD Mode Option RequiredAddressed commands with response Read registers Print Power On TimeRetransmit Last Message Test if MD Mode is Installed Addressed commands without response Acknowledge SRQRe-enable SRQ with out reading/clearing the Seve Register 110x STATUS, Error and SRQ Registers Status Control CommandsConditional Registers TXDService Request Enable and Event Registers BITFault Enable Register Fault Event RegisterStatus Enable Register Status Event RegisterPower supply set-up Serial Communication Test SET-UPCommunication Test 1 0-5V/0-10V Option PN IS510 SpecificationsIsolated Analog Programming Option PPM/CIsolated Programming & Monitoring Connector Setup and Operating Instructions Maintenance OUP Fuse Rating