Agilent 8753ES Specifications
Test-Port Output Characteristics7
Frequency characteristics |
|
Range | 30 kHz to 3 GHz |
Option 006 | 30 kHz to 6 GHz |
Resolution | 1 Hz |
Stability9 | ±7.5 ppm (0° to 55° C) |
| ±3 ppm/year |
Option 1D5 | ±0.05 ppm (0° to 55° C), |
| ±0.5 ppm/year |
Accuracy | ±10 ppm at 25° ±5° C |
Power range2 | |
Option 075/014 | |
Power sweep range | 25 dB (typically 31 dB) |
Resolution | 0.01 dB |
Level accuracy1,2,5 | ±1.0 dB |
Level linearity1,2,5 | |
| 5 dBm to 10 dBm6 ±0.5 dB |
Impedance9 | 50 Ω (nominal) |
30 kHz to 3 GHz | >16 dB RL (<1.38 SWR) |
3 GHz to 6 GHz | >14 dB RL (<1.50 SWR) |
Option 075 | 75 Ω (nominal) |
30 kHz to 3 GHz | >16 dB RL (<1.38 SWR) |
Spectral purity |
|
2nd harmonic3 | |
| |
Option 002 | |
3rd harmonic4 | |
| |
Option 002 | |
Nonharmonic spurious9 |
|
Mixer related | |
|
Test-Port Input Characteristics
Frequency range | 30 kHz to 3 GHz |
Option 006 | 30 kHz to 6 GHz |
Average noise level2,8 |
|
<3 GHz | |
<3 GHz | |
<3 GHz | |
3 to 6 GHz | |
3 to 6 GHz | |
3 to 6 GHz | |
Maximum input level | 10 dBm |
Damage level | 26 dBm or 35 VDC |
Impedance | 50 Ω (nominal) |
Option 075 | 75 Ω (nominal) |
Frequency response2,5 |
|
300 kHz to 3 GHz | ±1.0 dB |
3 GHz to 6 GHz | ±2.0 dB |
Internally generated harmonics (Option 002) | |
2nd harmonic3 | |
| |
| |
3rd harmonic4 | |
| |
| |
Harmonic measurement accuracy10 | |
300 kHz to 3 GHz | ±1.5 dB |
3 GHz to 6 GHz | ±3 dB (Option 006) |
Harmonic measurement dynamic range | |
Frequency offset mode11 |
|
Frequency range | 300 kHz to 3 GHz |
Option 006 | 300 kHz to 6 GHz |
| |
Power level |
|
300 kHz to 3 GHz | 0 to |
3 GHz to 6 GHz | 0 to |
LO spectral purity (typical) |
|
Maximum spurious input | |
Residual FM | <20 kHz |
LO frequency accuracy9 | ±1 MHz at nominal frequency |
1. | Relative to 0 dBm output power. |
2. | Typical below 300 kHz. |
3. | 16 MHz to 3 GHz. |
4. | 16 MHz to 2 GHz. |
5. | Typical from 2 to 3 GHz for instruments with Option 075. |
6. | +8 dBm maximum with Option 075, or Option 014. |
7. | Test performed on port 1 only. |
8. | Instruments with Option 075 are degraded 2 dB. |
9. | Typical performance. |
10. | 25° ±5° C. |
11. The Agilent 8753ES source characteristics and measurement accuracy in this | |
| mode are dependent on the stability of the external LO source. The RF source |
| tracks the LO to maintain a stable IF signal at the |
| Degradation in accuracy is negligible when using an Agilent 8642A/B, 8656B, |
10 | or E4432B RF signal generator as the LO source. |