Agilent Technologies 8753ET Agilent 8753ES Specifications, Test-Port Output Characteristics7

Page 10

Agilent 8753ES Specifications

Test-Port Output Characteristics7

Frequency characteristics

 

Range

30 kHz to 3 GHz

Option 006

30 kHz to 6 GHz

Resolution

1 Hz

Stability9

±7.5 ppm (0° to 55° C)

 

±3 ppm/year

Option 1D5

±0.05 ppm (0° to 55° C),

 

±0.5 ppm/year

Accuracy

±10 ppm at 25° ±5° C

Power range2

-85 to +10 dBm

Option 075/014

-85 to +8 dBm

Power sweep range

25 dB (typically 31 dB)

Resolution

0.01 dB

Level accuracy1,2,5

±1.0 dB

Level linearity1,2,5

-15 dBm to +5 dBm ±0.2 dB

 

5 dBm to 10 dBm6 ±0.5 dB

Impedance9

50 Ω (nominal)

30 kHz to 3 GHz

>16 dB RL (<1.38 SWR)

3 GHz to 6 GHz

>14 dB RL (<1.50 SWR)

Option 075

75 Ω (nominal)

30 kHz to 3 GHz

>16 dB RL (<1.38 SWR)

Spectral purity

 

2nd harmonic3

<-40 dBc at 0 dBm9

 

<-50 dBc at -10 dBm9

Option 002

<-25 dBc at 10 dBm6

3rd harmonic4

<-40 dBc at 0 dBm9

 

<-50 dBc at -10 dBm9

Option 002

<-25 dBc at 10 dBm6

Nonharmonic spurious9

 

Mixer related

<-30 dBc at 10 dBm6

 

<-55 dBc at -10 dBm

Test-Port Input Characteristics

Frequency range

30 kHz to 3 GHz

Option 006

30 kHz to 6 GHz

Average noise level2,8

 

<3 GHz

<-82 dBm (3 kHz BW)

<3 GHz

<-102 dBm (10 Hz BW)

<3 GHz

<-110 dBm (10 Hz BW) typical

3 to 6 GHz

<-77 dBm (3 kHz BW)

3 to 6 GHz

<-97 dBm (10 Hz BW)

3 to 6 GHz

<-105 dBm (10 Hz BW) typical

Maximum input level

10 dBm

Damage level

26 dBm or 35 VDC

Impedance

50 Ω (nominal)

Option 075

75 Ω (nominal)

Frequency response2,5

 

300 kHz to 3 GHz

±1.0 dB

3 GHz to 6 GHz

±2.0 dB

Internally generated harmonics (Option 002)

2nd harmonic3

<-15 dBc at +8 dBm

 

<-30 dBc at 0 dBm9

 

<-45 dBc at -15 dBm9

3rd harmonic4

<-30 dBc at +8 dBm

 

<-50 dBc at 0 dBm9

 

<-50 dBc at -15 dBm9

Harmonic measurement accuracy10

300 kHz to 3 GHz

±1.5 dB

3 GHz to 6 GHz

±3 dB (Option 006)

Harmonic measurement dynamic range

-40 dBc (output = -10 dBm, input = <-15 dBm)9

Frequency offset mode11

 

Frequency range

300 kHz to 3 GHz

Option 006

300 kHz to 6 GHz

R-channel input requirements

 

Power level

 

300 kHz to 3 GHz

0 to -35 dBm

3 GHz to 6 GHz

0 to -30 dBm

LO spectral purity (typical)

 

Maximum spurious input

<-25 dBc

Residual FM

<20 kHz

LO frequency accuracy9

±1 MHz at nominal frequency

1.

Relative to 0 dBm output power.

2.

Typical below 300 kHz.

3.

16 MHz to 3 GHz.

4.

16 MHz to 2 GHz.

5.

Typical from 2 to 3 GHz for instruments with Option 075.

6.

+8 dBm maximum with Option 075, or Option 014.

7.

Test performed on port 1 only.

8.

Instruments with Option 075 are degraded 2 dB.

9.

Typical performance.

10.

25° ±5° C.

11. The Agilent 8753ES source characteristics and measurement accuracy in this

 

mode are dependent on the stability of the external LO source. The RF source

 

tracks the LO to maintain a stable IF signal at the R-channel receiver input.

 

Degradation in accuracy is negligible when using an Agilent 8642A/B, 8656B,

10

or E4432B RF signal generator as the LO source.

Image 10
Contents Agilent 8753ET/ES 8753ET and 8753ES Network Analyzers, 30 kHz to 3 or 6 GHzTable of Contents Dynamic Range Measurement Port CharacteristicsMeasurement Uncertainty System dynamic range ConfigurationAgilent 8753ES 50-ohm systems Reflection uncertaintyType-N test ports Calibration kit 85032B Test-port cables 11857DCalibration kit 85033D Test-port cables 11857D Network analyzer Agilent 8753ES 75-Ohm SystemsCalibration kit 85036B Test-port cables 11857B Type-F test ports Calibration kit 85039B Test-port cables 11857BPart number Measurement Port CharacteristicsAgilent 8753ET 50-ohm systems KHz to 16 MHz 100 dB MHz to 3 GHz 110 dBTest-Port Output Characteristics7 Agilent 8753ES SpecificationsTest-Port Input Characteristics Magnitude Characteristics Phase CharacteristicsExternal source mode1 CW time sweep only Test Port Output Characteristics Agilent 8753ES Option 011 SpecificationsTest Port Input Characteristics Option External source mode4 Frequency offset mode3Channel input requirements1 Agilent 8753ET Specifications Reference level Range Display resolutionInput frequency accuracy requirement2 Stability2 KHz to 3 GHzMeasurement Agilent 8753ET/ES Supplemental CharacteristicsGroup Delay Characteristics Source ControlData Averaging Data Accuracy EnhancementDisplay Control Data Hardcopy Trace FunctionsStorage System CapabilitiesTime-domain Option Test SequencesRemote Programming Front Panel Connectors General CharacteristicsRear-Panel Connectors Options Cabinet Dimensions 8753ET/ESGeneral Conditions Agilent 85046A/B Specifications Test Set Specifications for OptionAgilent 85046A/B S-Parameter Test Sets Specifications Agilent 85047A S-Parameter Test SetAccessories Calibration KitsAgilent 85038 Family of Calibration Kits Agilent 11906 Family of Adapter KitsVerification Kits Test-Port Return CablesRF Limiter Ohm Accessory Kits85024A High-Frequency Probe Weight Net 3.5 kg 7.7 lb Agilent 8347A RF Amplifier