Calibration Kits

Vector accuracy enhancement procedures require that the systematic errors of the measurement sys- tem be characterized by measuring known devices (standards) on the system over the frequency range of interest. Return loss specifications or typical values are provided where available for the terminations and adapters.

85031B 7-mm calibration kit

Contains precision 7-mm standards used to cali- brate the 8753ES for measurement of devices with precision 7-mm connectors.

Includes

 

Part number

7 mm short/open circuit

85031-60001

7 mm 50-ohm terminations:

00909-60008

(two each)

 

 

Specifications for terminations

 

DC to 5 GHz

RL 52 dB

 

5 to 6 GHz

RL 46 dB

 

85032B 50-ohm type-N calibration kit

Contains precision 50-ohm type-N standards used to calibrate the 8753ET/ES and 50-ohm test sets for measurement of devices with 50-ohm type-N connectors. Precision phase-matched 7-mm to type-N adapters are included for accurate meas- urements of non-insertable devices. Option 001 deletes the 7-mm to type-N adapters, which are not needed with the 8753ET.

Includes

 

Part number

N-male 50-ohm termination

00909-60009

N-female 50-ohm termination

00909-60010

N-male short circuit

85032-60008

N-female short circuit

85032-60009

N-female open circuit

85032-60012

N-male open circuit

85032-60007

7-mm to N-male adapter

85054-60009

(two each)

 

 

7-mm to N-female adapter

85054-60001

(two each)

 

 

Specifications for terminations:

 

DC to 3 GHz

RL 49 dB

 

2 to 3 GHz

RL 46 dB

 

3 to 6 GHz

RL 40 dB

 

Typical adapter characteristics:

 

DC to 6 GHz

RL 30 dB

 

85033D 3.5-mm calibration kit

Contains a set of precision 3.5-mm standards to calibrate the 8753ES and 50-ohm test sets for the measurement of devices with precision 3.5-mm and SMA connectors. Precision phase-matched

7-mm to 3.5-mm adapters are included for accurate measurements of non-insertable devices. Option 001 deletes the 7-mm to type-N adapters.

Includes

Part number

3.5-mm-male 50-ohm termination

85033-60009

3.5-mm-female 50-ohm termination 85033-60010

3.5-mm-female short

85033-60014

3.5-mm-male short

85033-60013

3.5-mm-female open

85033-60012

3.5-mm-male open

85033-60011

7-mm to 3.5-mm female adapter

1250-1747

(two)

 

 

7-mm to 3.5-mm male adapter

1250-1746

(two)

 

 

Specifications for terminations:

 

DC to 1.3 GHz

RL 46 dB

 

1.3 to 3 GHz

RL 44 dB

 

3 to 6 GHz

RL 38 dB

 

Typical adapter characteristics:

 

DC to 6 GHz

RL 34 dB

 

85036B 75-ohm type-N calibration kit

Contains a set of precision 75-ohm type-N stan- dards to calibrate the 8753ES with a 75-ohm test set for measurement of devices with 75-ohm type-N connectors. Precision phased-matched adapters are included for accurate measurements of non- insertable devices.

Includes

 

Part number

N-male 75-ohm termination

00909-60019

N-female 75-ohm termination

00909-60020

N-female 75-ohm short

85036-60011

N-male 75-ohm short

85036-60012

N-female open

 

85032-20001

N-male open

 

85032-60007

N-male to N-male 75-ohm adapter

85036-60013

N-female to N-female 75-ohm

85036-60014

adapter

 

 

N-male to N-female 75-ohm

85036-60015

adapter

 

 

Specifications for terminations:

 

DC to 2 GHz

RL 46 dB

 

2 to 3 GHz

RL 40 dB

 

24

Image 24
Contents Agilent 8753ET/ES 8753ET and 8753ES Network Analyzers, 30 kHz to 3 or 6 GHzTable of Contents Measurement Port Characteristics Dynamic RangeMeasurement Uncertainty Configuration Agilent 8753ES 50-ohm systemsSystem dynamic range Reflection uncertaintyType-N test ports Calibration kit 85032B Test-port cables 11857DCalibration kit 85033D Test-port cables 11857D Agilent 8753ES 75-Ohm Systems Network analyzerCalibration kit 85036B Test-port cables 11857B Type-F test ports Calibration kit 85039B Test-port cables 11857BMeasurement Port Characteristics Agilent 8753ET 50-ohm systemsPart number KHz to 16 MHz 100 dB MHz to 3 GHz 110 dBAgilent 8753ES Specifications Test-Port Output Characteristics7Test-Port Input Characteristics Phase Characteristics Magnitude CharacteristicsExternal source mode1 CW time sweep only Agilent 8753ES Option 011 Specifications Test Port Output CharacteristicsTest Port Input Characteristics Option Frequency offset mode3 External source mode4Channel input requirements1 Agilent 8753ET Specifications Display resolution Input frequency accuracy requirement2Reference level Range Stability2 KHz to 3 GHzAgilent 8753ET/ES Supplemental Characteristics Group Delay CharacteristicsMeasurement Source ControlData Accuracy Enhancement Data AveragingDisplay Control Trace Functions StorageData Hardcopy System CapabilitiesTest Sequences Time-domain OptionRemote Programming General Characteristics Front Panel ConnectorsRear-Panel Connectors Cabinet Dimensions 8753ET/ES OptionsGeneral Conditions Test Set Specifications for Option Agilent 85046A/B SpecificationsAgilent 85046A/B S-Parameter Test Sets Specifications Agilent 85047A S-Parameter Test SetAccessories Calibration KitsAgilent 85038 Family of Calibration Kits Agilent 11906 Family of Adapter KitsVerification Kits Test-Port Return CablesOhm Accessory Kits RF Limiter85024A High-Frequency Probe Weight Net 3.5 kg 7.7 lb Agilent 8347A RF Amplifier