Agilent Technologies 8753ES, 8753ET manual Calibration kit 85033D Test-port cables 11857D

Page 6

System Performance Summaries

Agilent 8753ES (50-ohm systems)

3.5-mm test ports

Configuration

 

Network analyzer

8753ES

 

Standard and Option 006

Calibration kit

85033D

Test-port cables

11857D

System dynamic range

 

30 kHz to 50 kHz

70 dB1

50 kHz to 300 kHz

90 dB1

300 kHz to 16 MHz

100 dB

16 MHz to 3 GHz

110 dB

Option 014

108 dB

3 GHz to 6 GHz

105 dB

Option 014

103 dB

Reflection uncertainty

Magnitude

Measurement Port Characteristics

Frequency Range

Corrected

30 kHz–

300 kHz–

1.3 GHz–

3

GHz–

 

300 kHz2

1.3 GHz

3 GHz

6

GHz

 

 

 

 

 

Directivity

49 dB

46 dB

44 dB

38 dB

 

 

 

 

 

Source match

49 dB

44 dB

41 dB

37 dB

 

 

 

 

 

Load match

49 dB

46 dB

44 dB

38 dB

 

 

 

 

 

Reflection

±(0.010 dB

±(0.005 dB

±(0.007 dB

±(0.009 dB

tracking

+0.02/°C)

+0.01/°C)

+0.02/°C)

+0.03/°C)

 

 

 

 

 

Transmission

±(0.016 dB

±(0.014 dB

±(0.022 dB

±(0.048 dB

tracking

+0.02/°C)

+0.01/°C)

+0.02/°C)

+0.03/°C)

 

 

 

 

 

 

Phase

Transmission uncertainty

Magnitude

Phase

 

1.

Typical below 300 kHz.

6

2.

Typical performance.

Image 6
Contents Agilent 8753ET/ES 8753ET and 8753ES Network Analyzers, 30 kHz to 3 or 6 GHzTable of Contents Measurement Port Characteristics Dynamic RangeMeasurement Uncertainty System dynamic range ConfigurationAgilent 8753ES 50-ohm systems Reflection uncertaintyType-N test ports Calibration kit 85032B Test-port cables 11857DCalibration kit 85033D Test-port cables 11857D Agilent 8753ES 75-Ohm Systems Network analyzerCalibration kit 85036B Test-port cables 11857B Type-F test ports Calibration kit 85039B Test-port cables 11857BPart number Measurement Port CharacteristicsAgilent 8753ET 50-ohm systems KHz to 16 MHz 100 dB MHz to 3 GHz 110 dBAgilent 8753ES Specifications Test-Port Output Characteristics7Test-Port Input Characteristics Phase Characteristics Magnitude CharacteristicsExternal source mode1 CW time sweep only Agilent 8753ES Option 011 Specifications Test Port Output CharacteristicsTest Port Input Characteristics Option Frequency offset mode3 External source mode4Channel input requirements1 Agilent 8753ET Specifications Reference level Range Display resolutionInput frequency accuracy requirement2 Stability2 KHz to 3 GHzMeasurement Agilent 8753ET/ES Supplemental CharacteristicsGroup Delay Characteristics Source ControlData Accuracy Enhancement Data AveragingDisplay Control Data Hardcopy Trace FunctionsStorage System CapabilitiesTest Sequences Time-domain OptionRemote Programming General Characteristics Front Panel ConnectorsRear-Panel Connectors Cabinet Dimensions 8753ET/ES OptionsGeneral Conditions Test Set Specifications for Option Agilent 85046A/B SpecificationsAgilent 85046A/B S-Parameter Test Sets Specifications Agilent 85047A S-Parameter Test SetAccessories Calibration KitsAgilent 85038 Family of Calibration Kits Agilent 11906 Family of Adapter KitsVerification Kits Test-Port Return CablesOhm Accessory Kits RF Limiter85024A High-Frequency Probe Weight Net 3.5 kg 7.7 lb Agilent 8347A RF Amplifier
Related manuals