Test Set Specifications for Option 011
Agilent 85046A/B S-Parameter Test Sets
The 85046A/B
2 GHz. The 85046B has
Agilent 85046A/B Specifications
Note: Specifications that apply only to the 85046B are indicated in parentheses.
Impedance | 50 ohm (75 ohm) | |
Frequency range | 300 kHz to 3 GHz | |
|
| (300 kHz to 2 GHz) |
Directivity | 35 dB to 1.3 GHz | |
Typical tracking | 30 dB to Fmax1 | |
| ||
Transmission magnitude, phase2 | ||
0.3 | MHz to 2.0 MHz | ±1.5 dB, ±20° |
2.0 | MHz to Fmax | ±1.5 dB, ±10° |
Reflection magnitude, phase2 |
| |
0.3 | MHz to 2.0 MHz | ±1.5 dB, ±25° |
2.0 | MHz to Fmax | ±1.5 dB, ±10° |
Effective source match |
| |
0.3 MHz to 2.0 MHz | 14 dB | |
2.0 | MHz to 1.3 GHz | 20 dB (17 dB) |
1.3 | GHz to Fmax | 16 dB |
Nominal insertion loss
Input to test port
Test set switch/repeatability |
|
| ±0.03 dB; ±0.01 dB3 |
Max. operating level | +20 dBm |
Damage level | +30 dBm |
RF attenuator range | 70 dB, 10 dB steps |
DC bias range | ±30 VDC, 200 mA4 |
| 500 mA max |
DC bias connectors | 50 ohm BNC (f) |
Includes | four 190 mm (7.5 in) |
| cables and test set |
| interconnect cable. |
Dimensions | 90 mm H x 432 mm W |
| x 553 mm D |
Weight | 9.1 kg (20 lb) |
A standard 85046A/B test set contains a solid- state transfer switch, which allows continuous switching of power from port 1 to port 2 for full
Agilent 85046A Schematic
Input to incident
Port 1, 2 to A, B
| 1. | Fmax is the upper frequency limit of the associated test set. |
| 2. | Degrees, specified as deviation from linear phase. |
22 | 3. | Typical performance. |
4. | Some degradation of RF specifications may occur. |