Agilent Technologies 8753ET, 8753ES Input frequency accuracy requirement2, Display resolution

Page 15

External source mode1 (CW time sweep only)

Frequency range

300 kHz to 6 GHz

R-channel input requirements2

Power level

0 to -25 dBm

Spectral purity

 

Maximum spurious input

<-30 dBc

Residual FM

<20 kHz

Typical settling time

500 ms (automatic)

 

50 ms (manual)

Frequency readout accuracy2

0.1% (automatic)

Input frequency accuracy requirement2

Manual

-0.5 to 5 MHz

Accuracy

(See magnitude and phase

 

characteristics)

Magnitude characteristics

Dynamic accuracy

(10 Hz IF BW)

Display resolution

0.001 dB/division

Marker resolution3

0.001 dB

Trace noise

 

(0 dBm at transmission port or +5 dBm at reflection port, ratio measurement, 3 kHz BW)

300 kHz to 3 GHz

<0.006 dB rms

3 GHz to 6 GHz

<0.010 dB rms

1.See the Agilent 8753ET/ES descriptions and options for a functional description. Measurement accuracy is dependent on the stability of the input signal.

2.Typical performance.

3.Marker resolution for magnitude, phase and delay is dependent upon measured value. Resolution is limited to five digits.

Reference level

 

Range

±500 dB

Resolution

0.001 dB

Stability2

 

300 kHz to 3 GHz

0.02 dB/1° C

3 GHz to 6 GHz

0.04 dB/1° C

Phase Characteristics

Dynamic accuracy (10 Hz IF BW)

Range

±180°

Display resolution

0.01°/division

Marker resolution3

0.01°

Trace noise

 

(0 dBm at transmission port or +5 dBm at reflection

port, ratio measurement, 3 kHz BW)

300 kHz to 3 GHz

<0.038° rms

3 GHz to 6 GHz

<0.070° rms

Reference level

 

Range

-180° to +180°

Resolution

0.01°

Stability

 

30 kHz to 3 GHz

0.05°/° C

3 GHz to 6 GHz

0.20°/° C

Polar characteristics

 

Range

10 x 10–12to 1000 units full

 

scale

Reference

±500 units

15

Image 15 Contents
8753ET and 8753ES Network Analyzers, 30 kHz to 3 or 6 GHz Agilent 8753ET/ESTable of Contents Measurement Port Characteristics Dynamic RangeMeasurement Uncertainty Reflection uncertainty ConfigurationAgilent 8753ES 50-ohm systems System dynamic rangeCalibration kit 85032B Test-port cables 11857D Type-N test portsCalibration kit 85033D Test-port cables 11857D Agilent 8753ES 75-Ohm Systems Network analyzerCalibration kit 85036B Test-port cables 11857B Calibration kit 85039B Test-port cables 11857B Type-F test portsKHz to 16 MHz 100 dB MHz to 3 GHz 110 dB Measurement Port CharacteristicsAgilent 8753ET 50-ohm systems Part numberAgilent 8753ES Specifications Test-Port Output Characteristics7Test-Port Input Characteristics Phase Characteristics Magnitude CharacteristicsExternal source mode1 CW time sweep only Agilent 8753ES Option 011 Specifications Test Port Output CharacteristicsTest Port Input Characteristics Option Frequency offset mode3 External source mode4Channel input requirements1 Agilent 8753ET Specifications Stability2 KHz to 3 GHz Display resolutionInput frequency accuracy requirement2 Reference level RangeSource Control Agilent 8753ET/ES Supplemental CharacteristicsGroup Delay Characteristics MeasurementData Accuracy Enhancement Data AveragingDisplay Control System Capabilities Trace FunctionsStorage Data HardcopyTest Sequences Time-domain OptionRemote Programming General Characteristics Front Panel ConnectorsRear-Panel Connectors Cabinet Dimensions 8753ET/ES OptionsGeneral Conditions Test Set Specifications for Option Agilent 85046A/B SpecificationsAgilent 85046A/B S-Parameter Test Sets Agilent 85047A S-Parameter Test Set SpecificationsCalibration Kits AccessoriesAgilent 11906 Family of Adapter Kits Agilent 85038 Family of Calibration KitsTest-Port Return Cables Verification KitsOhm Accessory Kits RF Limiter85024A High-Frequency Probe Agilent 8347A RF Amplifier Weight Net 3.5 kg 7.7 lb