Agilent Technologies manual Agilent 8753ES 50-ohm systems, Configuration, System dynamic range

Page 4

System Performance Summaries

Agilent 8753ES (50-ohm systems)

7-mm test ports

Configuration

 

Network analyzer

8753ES

 

Standard and Option 006

Calibration kit

85031B

Test-port cables

11857D

System dynamic range

 

30 kHz to 50 kHz

70 dB1

50 kHz to 300 kHz

90 dB1

300 kHz to 16 MHz

100 dB

16 MHz to 3 GHz

110 dB

Option 014

108 dB

3 GHz to 6 GHz

105 dB

Option 014

103 dB

Reflection uncertainty

Magnitude

Transmission uncertainty

Magnitude

Measurement Port Characteristics

Frequency Range

Corrected

30 kHz–

300 kHz–

1.3 GHz–

3 GHz–

 

300 kHz2

1.3 GHz

3 GHz

6 GHz

Directivity

55 dB

55 dB

51 dB

46 dB

Source match

55 dB

51 dB

49 dB

43 dB

Load match

55 dB

55 dB

51 dB

46 dB

Reflection

±(0.001 dB

±(0.001 dB

±(0.005 dB

±(0.020 dB

tracking

+0.02 dB/°C)

+0.01 dB/°C)

+0.02 dB/°C)

+0.03 dB/°C)

Transmission

±(0.008 dB

±(0.006 dB

±(0.009 dB

±(0.021 dB

tracking

+0.02 dB/°C)

+0.01 dB/°C)

+0.02 dB/°C)

+0.03 dB/°C)

Uncorrected

 

 

 

 

Directivity

20 dB*

35 dB

30 dB

25 dB

Source match

18 dB**

16 dB

16 dB

14 dB

Load match

18 dB**

18 dB

16 dB

14 dB

Reflection

±2.5 dB

±1 dB

±1 dB

±1.5 dB

tracking

 

 

 

 

Transmission

±2.5 dB

±1 dB

±1 dB

±1.5 dB

tracking

 

 

 

 

Crosstalk

90 dB***

100 dB

100 dB

90 dB

*15 dB, 30 kHz to 50 kHz

**10 dB, 30 kHz to 50 kHz ***60 dB, 30 kHz to 50 kHz

Phase

Phase

4

1.

Typical below 300 kHz.

2.

Typical performance.

Image 4
Contents Agilent 8753ET/ES 8753ET and 8753ES Network Analyzers, 30 kHz to 3 or 6 GHzTable of Contents Dynamic Range Measurement Port CharacteristicsMeasurement Uncertainty Configuration Agilent 8753ES 50-ohm systemsSystem dynamic range Reflection uncertaintyType-N test ports Calibration kit 85032B Test-port cables 11857DCalibration kit 85033D Test-port cables 11857D Network analyzer Agilent 8753ES 75-Ohm SystemsCalibration kit 85036B Test-port cables 11857B Type-F test ports Calibration kit 85039B Test-port cables 11857BMeasurement Port Characteristics Agilent 8753ET 50-ohm systemsPart number KHz to 16 MHz 100 dB MHz to 3 GHz 110 dBTest-Port Output Characteristics7 Agilent 8753ES SpecificationsTest-Port Input Characteristics Magnitude Characteristics Phase CharacteristicsExternal source mode1 CW time sweep only Test Port Output Characteristics Agilent 8753ES Option 011 SpecificationsTest Port Input Characteristics Option External source mode4 Frequency offset mode3Channel input requirements1 Agilent 8753ET Specifications Display resolution Input frequency accuracy requirement2Reference level Range Stability2 KHz to 3 GHzAgilent 8753ET/ES Supplemental Characteristics Group Delay CharacteristicsMeasurement Source ControlData Averaging Data Accuracy EnhancementDisplay Control Trace Functions StorageData Hardcopy System CapabilitiesTime-domain Option Test SequencesRemote Programming Front Panel Connectors General CharacteristicsRear-Panel Connectors Options Cabinet Dimensions 8753ET/ESGeneral Conditions Agilent 85046A/B Specifications Test Set Specifications for OptionAgilent 85046A/B S-Parameter Test Sets Specifications Agilent 85047A S-Parameter Test SetAccessories Calibration KitsAgilent 85038 Family of Calibration Kits Agilent 11906 Family of Adapter KitsVerification Kits Test-Port Return CablesRF Limiter Ohm Accessory Kits85024A High-Frequency Probe Weight Net 3.5 kg 7.7 lb Agilent 8347A RF Amplifier