Agilent Technologies 8753ES manual Agilent 8753ET Specifications

Page 14

Agilent 8753ET Specifications

Test-Port Output Characteristics

Test Port Input Characteristics

Frequency characteristics Range

Option 006

Resolution

Stability4

Option 1D5

Accuracy

Power range

Resolution

Level accuracy1

Level linearity1

Standard

-20 to -15 dBm -15 to 0 dBm 0 to +5 dBm

Option 004

-15 to +5 dBm +5 to +10 dBm

Impedance

300 kHz to 3 GHz

3 GHz to 6 GHz

Spectral purity 2nd harmonic2

Option 002 3rd harmonic3

Option 002

Nonharmonic spurious

300 kHz to 3 GHz

300 kHz to 6 GHz

1 Hz

±7.5 ppm (0° to 55° C) ±3 ppm/year

±0.05 ppm (0° to 55° C) ±0.5 ppm/year

±10 ppm at 25° ±5° C -20 to +5 dBm

-85 to +10 dBm (with Option 004)

0.01dB ±1.0 dB

±0.5 dB ±0.2 dB ±0.5 dB

±0.2 dB ±0.5 dB

50 Ω (nominal)

>18 dB RL (< 1.28 SWR) >14 dB RL (<1.50 SWR)

<-40 dBc at 0 dBm4 <-50 dBc at -10 dBm4

<-25 dBc at maximum output power

<-40 dBc at 0 dBm <-50 dBc at -10 dBm

<-25 dBc at maximum output power

Frequency range

300 kHz to 3 GHz

Option 006

300 kHz to 6 GHz

Average noise level

 

300 kHz to 3 GHz

<-90 dBm (3 kHz BW)

300 kHz to 3 GHz

<-110 dBm (10 Hz BW)

300 kHz to 3 GHz

<-120 dBm (10 Hz BW)

 

typical

3 GHz to 6 GHz

<-85 dBm (3 kHz BW)

3 GHz to 6 GHz

<-105 dBm (10 Hz BW)

3 GHz to 6 GHz

<-114 dBm (10 Hz BW)

 

typical

Maximum input level

+10 dBm reflection port

 

0 dBm transmission port

Damage level

>+26 dBm or 35 VDC

Impedance

50 Ω (nominal)

Frequency response5

 

300 kHz to 3 GHz

±1.0 dB

3 GHz to 6 GHz

±2.0 dB

Harmonics (Option 002)

 

2nd harmonic2

<-15 dBc at +8 dBm

 

<-30 dBc at 0 dBm4

 

<-45 dBc at -15 dBm4

3rd harmonic3

<-30 dBc at +8 dBm

 

<-50 dBc at 0 dBm4

 

<-50 dBc at -15 dBm4

Harmonic measurement accuracy5

300 kHz to 3 GHz

±1.5 dB

3 GHz to 6 GHz

±3 dB (Option 006)

Harmonic measurement dynamic range4

-40 dBc (output = -10 dBm, input = <-15 dBm)

Frequency offset mode6

 

Frequency range

300 kHz to 3 GHz

Option 006

300 kHz to 6 GHz

R-channel input requirements

 

Power level

 

300 kHz to 3 GHz

0 to -35 dBm

3 GHz to 6 GHz

0 to -30 dBm

Mixer related

<-30 dBc at +10 dBm

 

<-55 dBc at -10 dBm

LO Spectral purity

 

Maximum spurious input

<-25 dBc

Residual FM

<20 kHz

LO frequency accuracy6

±1 MHz at nominal frequency

1.

Relative to -5 dBm output power for the Agilent 8753ET; -10 dBm output power

 

for 8753ET with Option 004.

2.

16 MHz to 3 GHz.

3.

16 MHz to 2 GHz.

4.

Typical performance.

5.

25° C ± 5° C.

6.

The Agilent 8753ET source characteristics and measurement accuracy in this

 

mode are dependent on the stability of the external LO source. The RF source

 

tracks the LO to maintain a stable IF signal at the R-channel receiver input.

14

Degradation in accuracy is negligible when using an Agilent 8642A/B, 8656B,

or E4432B RF signal generator as the LO source.

Image 14 Contents
Agilent 8753ET/ES 8753ET and 8753ES Network Analyzers, 30 kHz to 3 or 6 GHzTable of Contents Measurement Uncertainty Measurement Port CharacteristicsDynamic Range System dynamic range ConfigurationAgilent 8753ES 50-ohm systems Reflection uncertaintyType-N test ports Calibration kit 85032B Test-port cables 11857DCalibration kit 85033D Test-port cables 11857D Calibration kit 85036B Test-port cables 11857B Agilent 8753ES 75-Ohm SystemsNetwork analyzer Type-F test ports Calibration kit 85039B Test-port cables 11857BPart number Measurement Port CharacteristicsAgilent 8753ET 50-ohm systems KHz to 16 MHz 100 dB MHz to 3 GHz 110 dBTest-Port Input Characteristics Agilent 8753ES SpecificationsTest-Port Output Characteristics7 External source mode1 CW time sweep only Phase CharacteristicsMagnitude Characteristics Test Port Input Characteristics Option Agilent 8753ES Option 011 SpecificationsTest Port Output Characteristics Channel input requirements1 Frequency offset mode3External source mode4 Agilent 8753ET Specifications Reference level Range Display resolutionInput frequency accuracy requirement2 Stability2 KHz to 3 GHzMeasurement Agilent 8753ET/ES Supplemental CharacteristicsGroup Delay Characteristics Source ControlDisplay Control Data Accuracy EnhancementData Averaging Data Hardcopy Trace FunctionsStorage System CapabilitiesRemote Programming Test SequencesTime-domain Option Rear-Panel Connectors General CharacteristicsFront Panel Connectors General Conditions Cabinet Dimensions 8753ET/ESOptions Agilent 85046A/B S-Parameter Test Sets Test Set Specifications for OptionAgilent 85046A/B Specifications Specifications Agilent 85047A S-Parameter Test SetAccessories Calibration KitsAgilent 85038 Family of Calibration Kits Agilent 11906 Family of Adapter KitsVerification Kits Test-Port Return Cables85024A High-Frequency Probe Ohm Accessory KitsRF Limiter Weight Net 3.5 kg 7.7 lb Agilent 8347A RF Amplifier