Agilent Technologies 8753ET, 8753ES manual Agilent 85047A S-Parameter Test Set, Specifications

Page 23

Agilent 85047A S-Parameter Test Set

The 85047A S-parameter test set provides the capability to simultaneously measure the reflec- tion and transmission characteristics of two-port devices in either direction with a single connec- tion. This test set includes a frequency doubler that can be switched in by an 8753B/C Option 006 to measure 3 MHz to 6 GHz in a single sweep or switched out to measure 300 kHz to 3 GHz in

a single sweep. The 8753ES Option 011 does not use the frequency doubler, so the full 300 kHz to 6 GHz range is available. This test set exhibits

<5 dB insertion loss between the RF input and the test ports for as high as 15 dBm at the test port, and also includes a programmable step attenuator. There are two rear panel BNC outputs. One pro- vides a TTL signal which indicates the result of

a limit test. The second TTL output is controlled from the 8753ES test sequence function.

Specifications

Impedance

50 ohms

Frequency range

300 kHz to 3 GHz and

 

3 GHz to 6 GHz

 

(8753B/C);

 

300 kHz to 6 GHz

 

(8753D/E/ES Opt. 006)

Directivity1

 

300 kHz to 1.3 GHz

35 dB2

1.3 GHz to 3 GHz

30 dB

3 GHz to 6 GHz

25 dB

Typical tracking1

 

Transmission magnitude, phase3

300 kHz to 3 GHz

±1.5 dB, ±10°

3 GHz to 6 GHz

+0.5, -2.5 dB, ±20°

Reflection magnitude, phase3

 

300 kHz to 3 GHz

±1.5 dB, ±10°

3 GHz to 6 GHz

±1.5 dB, ±20°

Source match1

 

300 kHz to 1.3 GHz

20 dB

1.3 GHz to 3 GHz

16 dB

3 GHz to 6 GHz

14 dB

Normal insertion loss

 

Input to port 1,2

 

300 kHz to 3 GHz

4.0 dB +0.8 dB/GHz

3 GHz to 6 GHz

17.5 dB +0.8 dB/GHz

Input to R

 

300 kHz to 3 GHz

19 dB +0.5 dB/GHz

3 GHz to 6 GHz

34 dB +0.5 dB/GHz

Port 1,2 to A,B

16 dB

Typical isolation4

 

300 kHz to 3 GHz

100 dB

3 GHz to 6 GHz

90 dB

Test port switch

±0.03 dB; ±0.01 dB4

repeatability

 

Maximum operating level

+20 dBm

Damage level

+30 dBm

RF attenuator range

70 dB (10 dB steps)

DC bias range

±30 VDC, 200 mA,

 

no degradation in RF specs,

 

1A max.

RF connectors

 

Port 1,2

7 mm precision

All others

50 ohm type-N(f)

Dimensions

90 mm H x 432 mm W x

 

553 mm D

Weight

10 kg (22 lb)

A standard 85047A test set contains a solid-state transfer switch, which allows continuous switching of power from port 1 to port 2 for full two-port error correction.

Agilent 85047A Schematic

1.This can be greatly improved with accuracy enhancement.

2.Some degradation at environmental extremes below 600 kHz.

3.Degrees, specified as deviation from linear phase.

4. Typical performance.

23

Image 23 Contents
8753ET and 8753ES Network Analyzers, 30 kHz to 3 or 6 GHz Agilent 8753ET/ESTable of Contents Measurement Uncertainty Measurement Port CharacteristicsDynamic Range Reflection uncertainty ConfigurationAgilent 8753ES 50-ohm systems System dynamic rangeCalibration kit 85032B Test-port cables 11857D Type-N test portsCalibration kit 85033D Test-port cables 11857D Calibration kit 85036B Test-port cables 11857B Agilent 8753ES 75-Ohm SystemsNetwork analyzer Calibration kit 85039B Test-port cables 11857B Type-F test portsKHz to 16 MHz 100 dB MHz to 3 GHz 110 dB Measurement Port CharacteristicsAgilent 8753ET 50-ohm systems Part numberTest-Port Input Characteristics Agilent 8753ES SpecificationsTest-Port Output Characteristics7 External source mode1 CW time sweep only Phase CharacteristicsMagnitude Characteristics Test Port Input Characteristics Option Agilent 8753ES Option 011 SpecificationsTest Port Output Characteristics Channel input requirements1 Frequency offset mode3External source mode4 Agilent 8753ET Specifications Stability2 KHz to 3 GHz Display resolutionInput frequency accuracy requirement2 Reference level RangeSource Control Agilent 8753ET/ES Supplemental CharacteristicsGroup Delay Characteristics MeasurementDisplay Control Data Accuracy EnhancementData Averaging System Capabilities Trace FunctionsStorage Data HardcopyRemote Programming Test SequencesTime-domain Option Rear-Panel Connectors General CharacteristicsFront Panel Connectors General Conditions Cabinet Dimensions 8753ET/ESOptions Agilent 85046A/B S-Parameter Test Sets Test Set Specifications for OptionAgilent 85046A/B Specifications Agilent 85047A S-Parameter Test Set SpecificationsCalibration Kits AccessoriesAgilent 11906 Family of Adapter Kits Agilent 85038 Family of Calibration KitsTest-Port Return Cables Verification Kits85024A High-Frequency Probe Ohm Accessory KitsRF Limiter Agilent 8347A RF Amplifier Weight Net 3.5 kg 7.7 lb