Agilent Technologies 8753ET manual Agilent 8753ES Option 011 Specifications

Page 12

Agilent 8753ES Option 011 Specifications

Test Port Output Characteristics

Frequency characteristics

 

Range

300 kHz to 3 GHz

Option 006

30 kHz to 6 GHz

Resolution

1 Hz

Stability

typically ±7.5 ppm 0° to 55° C

 

typically ±3 ppm/year

Option 1D5

typically ±0.05 ppm 0° to 55° C

 

typically ±0.5 ppm/year

Accuracy

±10 ppm at 25° ± 5° C

Power range

-5 to +20 dBm

Option 006

-5 to +18 dBm

Resolution

0.01 dB

Level accuracy1,2

±1.0 dB

Level linearity1,2,5

±0.25 dB, -5 to +15 dBm

 

±0.5 dB, +15 to +20 dBm

Impedance

50 Ω nominal

300 kHz to 3 GHz2 >16 dB RL (<1.38 SWR)

300 kHz to 6 GHz

>14 dB RL (<1.50 SWR)

Spectral purity

 

2nd harmonic3

<-40 dBc at +10 dBm6

 

<-50 dBc at 0 dBm6

Option 002

<-25 dBc at max power

3rd harmonic4

<-40 dBc at +10 dBm6

 

<-50 dBc at 0 dBm6

Option 002

<-25 dBc at max power

Nonharmonic spurious6

 

Mixer related

<-30 dBc at max power

 

<-55 dBc at -10 dBm

Test Port Input Characteristics Option 011

Frequency range

300 kHz to 3 GHz

Option 006

30 kHz to 6 GHz

Average noise level2

 

50 kHz to 3 GHz

<-90 dBm (3 kHz BW)

50 kHz to 3 GHz

<-110 dBm (10 Hz BW)

50 kHz to 3 GHz

<-120 dBm (10 Hz BW) typical

3 to 6 GHz

<-85 dBm (3 kHz BW)

3 to 6 GHz

<-105 dBm (10 Hz BW)

3 to 6 GHz

<-115 dBm (10 Hz BW) typical

Maximum input level

0 dBm

Damage level

20 dBm or 25 VDC

Impedance

50 ohms nominal

300 kHz to 2 MHz

≥20 dB RL

2 MHz to 1.3 GHz

≥24 dB RL

1.3 GHz to 3 GHz

≥19 dB RL

Option 006

≥15 dB RL6

3 GHz to 6 GHz

Frequency response

 

300 kHz to 3 GHz

±1.0 dB

3 GHz to 6 GHz

±2.0 dB

Harmonics (Option 002)

 

2nd harmonic3

<-15 dBc at 0 dBm

 

<-30 dBc at -10 dBm6

 

<-45 dBc at -30 dBm6

3rd harmonic4

<-30 dBc at 0 dBm

 

<-50 dBc at -10 dBm6

 

<-50 dBc at -30 dBm6

Harmonic measurement accuracy

16 MHz to 3 GHz

±1.5 dB

3 GHz to 6 GHz Option 006

±3 dB

Harmonic measurement dynamic range

-40 dBc6 (output = -10 dBm, input <-15 dBm)

 

1.

+10 dBm output power for Agilent 8753ES Option 011.

 

2.

Typical below 300 kHz.

 

3.

16 MHz to 3 GHz.

 

4.

16 MHz to 2 GHz.

 

5.

For Agilent 8753ES Option 011 and Option 006, linearity is specified for the

12

 

ranges of -5 to +13 dBm and +13 to +18 dBm.

6.

Typical performance.

Image 12
Contents Agilent 8753ET/ES 8753ET and 8753ES Network Analyzers, 30 kHz to 3 or 6 GHzTable of Contents Measurement Port Characteristics Dynamic RangeMeasurement Uncertainty Configuration Agilent 8753ES 50-ohm systemsSystem dynamic range Reflection uncertaintyType-N test ports Calibration kit 85032B Test-port cables 11857DCalibration kit 85033D Test-port cables 11857D Agilent 8753ES 75-Ohm Systems Network analyzerCalibration kit 85036B Test-port cables 11857B Type-F test ports Calibration kit 85039B Test-port cables 11857BMeasurement Port Characteristics Agilent 8753ET 50-ohm systemsPart number KHz to 16 MHz 100 dB MHz to 3 GHz 110 dBAgilent 8753ES Specifications Test-Port Output Characteristics7Test-Port Input Characteristics Phase Characteristics Magnitude CharacteristicsExternal source mode1 CW time sweep only Agilent 8753ES Option 011 Specifications Test Port Output CharacteristicsTest Port Input Characteristics Option Frequency offset mode3 External source mode4Channel input requirements1 Agilent 8753ET Specifications Display resolution Input frequency accuracy requirement2Reference level Range Stability2 KHz to 3 GHzAgilent 8753ET/ES Supplemental Characteristics Group Delay CharacteristicsMeasurement Source ControlData Accuracy Enhancement Data AveragingDisplay Control Trace Functions StorageData Hardcopy System CapabilitiesTest Sequences Time-domain OptionRemote Programming General Characteristics Front Panel ConnectorsRear-Panel Connectors Cabinet Dimensions 8753ET/ES OptionsGeneral Conditions Test Set Specifications for Option Agilent 85046A/B SpecificationsAgilent 85046A/B S-Parameter Test Sets Specifications Agilent 85047A S-Parameter Test SetAccessories Calibration KitsAgilent 85038 Family of Calibration Kits Agilent 11906 Family of Adapter KitsVerification Kits Test-Port Return CablesOhm Accessory Kits RF Limiter85024A High-Frequency Probe Weight Net 3.5 kg 7.7 lb Agilent 8347A RF Amplifier