Agilent Technologies 8753ES, 8753ET manual Verification Kits, Test-Port Return Cables

Page 26

Accessories

Verification Kits

Measuring known devices other than the standards used in calibration is an easy way to verify the proper operation of an 8753ES measurement sys- tem. Agilent offers verification kits which include devices, with data, for verifying the error-corrected measurements of an 8753ES and 50-ohm test sets.

85029B 7-mm verification kit

Contains a set of precision 7-mm devices, with data traceable to NIST* for verifying the calibrated performance of an 8753E measurement system. The 85031B 7-mm calibration kit is required for complete verification.

Test-Port Return Cables

Agilent offers high-quality RF cables to connect the 8753ET/ES and test sets to devices under test. These cables offer excellent RF shielding for high- dynamic-range measurements.

11851B 50-ohm type-N RF cable kit

Recommended for use with the 11850C/D three-way power splitters. Kit includes three phase-matched 610-mm (24-in.) cables and one 860-mm (34-in.) cable.

Return loss

>24 dB to 3 GHz

Phase tracking

±4° at 1.3 GHz

8120-5639 50-ohm type-N RF cable A 50-ohmtype-Ntest-port cable (both connectors are type-N male).

Return loss>30dB to 6 GHz

11857B 75-ohm type-N test-port return cables

A pair of 610-mm (24-in.) test-port return cables for use with the 8753ES or 85046B 75-ohm

S-parameter test set.

 

Return loss

>24 dB to 2 GHz

Phase tracking

±2° at 1.3 GHz

11857D 7-mm test-port return cables

A pair of 610-mm (24-in.) test-port return cables for use with the 8753ES with the 85046A and 85047A S-parameter test sets. These cables can be used with connector types other than 7-mm with the appropriate precision adapters.

Return loss

>24 dB to 3 GHz

 

>20 dB to 6 GHz

Phase tracking

±2° at 1.3 GHz

11857F type-F test-port return cables

Type-F test-port cable set (75-ohm). Includes one cable with type-N(male) to type-F(male) connectors and one cable with type-N(male) to type-F(female) connectors.

11850C/D three-way power splitters

 

 

11850C

 

11850D

Impedance

50 ohms

 

75 ohms

Frequency range

DC to 3 GHz

DC to 2 GHz

Tracking

±25 dB, ±3°

±2 dB, ±2.5°

Equivalent source

30 dB at 1.3 GHz

30 dB at 1.3 GHz

match

 

 

 

(ratio or leveling)

20 dB at 3 GHz

20 dB at 2 GHz

Nominal

9.5 dB + 1 dB/GHz

7.8 dB

insertion loss

 

 

 

Input port match

 

 

 

DC to 1.3 GHz

20 dB

 

20 dB

1.3 GHz to Fmax

10 dB

 

10 dB

Maximum operating

+20 dB

 

+20 dB

level

 

 

 

Damage level

+30 dB

 

+30 dB

RF connectors

 

 

 

RF input

50-ohm type-N (f)

50-ohm type-N(f)

All others

50-ohm type-N (f)

75-ohm type-N(f)

Includes

Three 11852B 50- to 75-ohm

 

minimum-loss pads

 

Recommended

11851B RF cable kit

 

accessories

 

 

 

11667A 50-ohm power splitter

 

Frequency range

 

DC to 18 GHz

Typical insertion loss

 

6 dB

 

Equivalent source match

 

 

DC to 4 GHz

 

26 dB

 

4 GHz to 8 GHz

 

21 dB

 

8 GHz to 18 GHz

 

17 dB

 

Tracking (between output arms)

 

DC to 4 GHz

 

±15 dB

 

4 GHz to 8 GHz

 

±2 dB

 

8 GHz to 18 GHz

 

±25 dB

 

Maximum operating level

±27 dBm

 

Connectors

 

50-ohm type-N (f)

Option 001

 

 

 

RF input

 

type-N (m)

 

Outputs

 

type-N (f)

 

Option 002

 

 

 

RF input

 

type-N (f) on RF

Outputs

 

precision 7-mm

Dimensions

 

46 mmH x 52 mm W x19 mm D

 

 

(1.8 x 2.0 x 0.7 in)

Recommended accessories

11851B RF cable kit

11852B 50-ohm to 75-ohm minimum loss pad

Frequency range

 

DC to 3.0 GHz

Nominal insertion loss

5.7 dB

 

Return loss

 

 

 

300 kHz to 2 GHz

 

32 dB

 

2 GHz to 3 GHz

 

27 dB

 

Maximum input power

250 mW (+24 dBm)

26

Image 26 Contents
Agilent 8753ET/ES 8753ET and 8753ES Network Analyzers, 30 kHz to 3 or 6 GHzTable of Contents Measurement Uncertainty Measurement Port CharacteristicsDynamic Range System dynamic range ConfigurationAgilent 8753ES 50-ohm systems Reflection uncertaintyType-N test ports Calibration kit 85032B Test-port cables 11857DCalibration kit 85033D Test-port cables 11857D Calibration kit 85036B Test-port cables 11857B Agilent 8753ES 75-Ohm SystemsNetwork analyzer Type-F test ports Calibration kit 85039B Test-port cables 11857BPart number Measurement Port CharacteristicsAgilent 8753ET 50-ohm systems KHz to 16 MHz 100 dB MHz to 3 GHz 110 dBTest-Port Input Characteristics Agilent 8753ES SpecificationsTest-Port Output Characteristics7 External source mode1 CW time sweep only Phase CharacteristicsMagnitude Characteristics Test Port Input Characteristics Option Agilent 8753ES Option 011 SpecificationsTest Port Output Characteristics Channel input requirements1 Frequency offset mode3External source mode4 Agilent 8753ET Specifications Reference level Range Display resolutionInput frequency accuracy requirement2 Stability2 KHz to 3 GHzMeasurement Agilent 8753ET/ES Supplemental CharacteristicsGroup Delay Characteristics Source ControlDisplay Control Data Accuracy EnhancementData Averaging Data Hardcopy Trace FunctionsStorage System CapabilitiesRemote Programming Test SequencesTime-domain Option Rear-Panel Connectors General CharacteristicsFront Panel Connectors General Conditions Cabinet Dimensions 8753ET/ESOptions Agilent 85046A/B S-Parameter Test Sets Test Set Specifications for OptionAgilent 85046A/B Specifications Specifications Agilent 85047A S-Parameter Test SetAccessories Calibration KitsAgilent 85038 Family of Calibration Kits Agilent 11906 Family of Adapter KitsVerification Kits Test-Port Return Cables85024A High-Frequency Probe Ohm Accessory KitsRF Limiter Weight Net 3.5 kg 7.7 lb Agilent 8347A RF Amplifier