Agilent 8753ES (50-ohm systems)
Type-N test ports
Configuration |
|
Network analyzer | 8753ES |
| Standard and Option 006 |
Calibration kit | 85032B |
| 11857D |
System dynamic range |
|
30 kHz to 50 kHz | 70 dB1 |
50 kHz to 300 kHz | 90 dB1 |
300 kHz to 16 MHz | 100 dB |
16 MHz to 3 GHz | 110 dB |
Option 014 | 108 dB |
3 GHz to 6 GHz | 105 dB |
Option 014 | 103 dB |
Reflection uncertainty
Magnitude
Transmission uncertainty
Magnitude
Measurement Port Characteristics
Frequency Range
Corrected | 30 kHz– | 300 kHz– | 1.3 GHz– | 3 | GHz– |
| 300 kHz2 | 1.3 GHz | 3 GHz | 6 | GHz |
|
|
|
|
| |
Directivity | 50 dB | 50 dB | 47 dB | 40 dB | |
|
|
|
|
| |
Source match | 49 dB | 42 dB | 36 dB | 31 dB | |
|
|
|
|
| |
Load match | 50 dB | 50 dB | 47 dB | 40 dB | |
|
|
|
|
| |
Reflection | ±(0.005 dB | ±(0.009 dB | ±(0.019 dB | ±(0.070 dB | |
tracking | +0.02 dB/°C) | +0.01 dB/°C) | +0.02 dB/°C) | +0.03 dB/°C) | |
|
|
|
|
| |
Transmission | ±(0.014 dB | ±(0.013 dB | ±(0.026 dB | ±(0.065 dB | |
tracking | +0.02 dB/°C) | +0.01 dB/°C) | +0.02 dB/°C) | +0.03 dB/°C) | |
|
|
|
|
|
|
Phase
Phase
1. | Typical below 300 kHz. | 5 |
2. | Typical performance. |