Sweep trigger
Set to either continuous, single, group sweep, or external trigger. Set external trigger to take a com- plete sweep or to measure individual points in a frequency, power or list sweep.
Power
Set source power from
Power Meter Calibration
Select continuous leveling or use a correction table to modify source power. The correction table is created with an initial single sweep. Make single or multiple power meter readings at each frequency.
Data Accuracy Enhancement
Measurement calibration
Measurement calibration significantly reduces measurement uncertainty due to errors caused by system directivity, source and load match, tracking, and crosstalk. Full
Calibration types available
•Frequency response
Simultaneous magnitude and phase correction of frequency response errors for either reflec- tion or transmission measurements.
•Response and isolation
Compensates for frequency response and direc- tivity (reflection) or frequency response and crosstalk errors.
•Enhanced response calibration
Corrects for frequency response and source match for transmission measurements, and provides
•One-port calibration
Uses test set port 1 or port 2 to correct for directivity, frequency response and source match errors.
•Two-port calibration
Compensates for directivity, source match, reflection frequency response, load match, transmission frequency response and crosstalk for an
•TRL*/LRM* calibration
Compensates for directivity, reflection and transmission frequency response, and crosstalk in both the forward and reverse directions.
Especially suitable for calibrating non- coaxial environments, such as in test fixtures. TRL*/LRM* is a special implementation of TRL/LRM calibration, modified for the three- sampler receiver in the 8753ES. Available on Agilent 8753ES analyzers.
Interpolated error correction
With any type of accuracy enhancement applied, interpolated mode recalculates the error coeffi- cients when the test frequencies are changed. The number of points can be increased or decreased and the start/stop frequencies can be changed, but the resulting frequency span must be a subset of the original calibration frequency span. System performance is not specified for measurements with interpolated error correction applied.
Velocity factor
Enters the velocity factor to calculate equivalent electrical length.
Reference plane extension
Redefine the
Select default calibration kit
Select from a list of standard calibration kits or choose a
Data Averaging
IF bandwidth
The IF bandwidth is selectable from 6 kHz to 10 Hz.
Weighted
Averages vector data on each successive sweep.
Trace smoothing
Computes the moving average of adjacent data points. Smoothing aperture defines the trace width (number of points) to be averaged, and ranges from 0.25% to 20% of the trace width.
Display Control
Display formats