General Characteristics
Front Panel Connectors
8753ES test ports (without Option 011)
Connector type | 7 mm, precision |
Impedance | 50 ohms (nominal) |
8753ES Option 011 test ports |
|
Connector type | |
Impedance | 50 ohms (nominal) |
8753ES Option 075 test ports |
|
Connector type | |
Impedance | 75 ohms (nominal) |
8753ET test ports |
|
Connector type | |
Impedance | 50 ohms (nominal) |
Probe power
+15V ±2% 400 mA
(combined load for both probe connections)
(combined load for both probe connections)
Rear-Panel Connectors
External reference frequency input (EXT REF INPUT)
Frequency | 1, 2, 5, and 10 MHz |
| (±200 Hz at 10 MHz) |
Level | |
Impedance | 50 ohms |
Connector | BNC (f) |
High-stability frequency reference output (Option 1D5)
Frequency | 10.0000 MHz |
Frequency stability | ±0.05 ppm |
(0° to 55° C) |
|
Daily aging rate | |
(after 30 days) |
|
Yearly aging rate | 0.5 ppm/year |
Output | 0 dBm minimum |
Nominal output impedance | 50 Ω Connector BNC (f) |
External auxiliary input (AUX INPUT) | |
Input voltage limits |
External AM input (EXT AM)
±1 volt into a 5 k Ω resistor, 1 kHz maximum,
resulting in approximately 8 dB/volt amplitude modulation. BNC (f) connector.
External trigger (EXT TRIGGER)
Triggers on a negative TTL transition or contact closure to ground. BNC (f) connector.
20
Test sequence output (TEST SEQ)
By default, this connector outputs a TTL
Limit test output (LIMIT TEST)
This connector outputs a TTL signal of the limit test results. Pass: TTL high. Fail: TTL low. BNC (f) connector.
Test-port bias input (BIAS CONNECT) (8753ES only)
Maximum voltage | +30 VDC |
Maximum current | ±200 mA |
(no degradation in RF specs) |
|
Maximum current | ±1 A |
Connector | BNC (f) |
VGA video output (EXT MON)
This connector drives external VGA monitors.
GPIB
This connector allows communications with compatible devices including external controllers, printers, plotters, disk drives, and power meters.
Parallel port
This
This
DIN keyboard
This
Test set interconnect
This connector is used to connect an 8753ES Option 011 to the 85046A/B or 85047A test set. 8753ES analyzers without Option 011 can use signal levels on this connector for sequencing or general purpose I/O applications.
Internal memory
Typical data retention time with 3V, 1.2 Ah battery:
At 25° C | 11,904 days (32.6 years) |
At 40° C | 1244 days (3.4 years) |
At 70° C | 250 days (0.68 year) |
Line power
48 Hz to 66 Hz
115V nominal (90V to 132V) or 230V nominal (198V to 264V). 280 VA max.