Agilent Technologies 8753ES, 8753ET manual General Characteristics, Front Panel Connectors

Page 20

General Characteristics

Front Panel Connectors

8753ES test ports (without Option 011)

Connector type

7 mm, precision

Impedance

50 ohms (nominal)

8753ES Option 011 test ports

 

Connector type

Type-N female

Impedance

50 ohms (nominal)

8753ES Option 075 test ports

 

Connector type

Type-N female

Impedance

75 ohms (nominal)

8753ET test ports

 

Connector type

Type-N female

Impedance

50 ohms (nominal)

Probe power

+15V ±2% 400 mA

(combined load for both probe connections) -12.6V ±5.5% 300 mA

(combined load for both probe connections)

Rear-Panel Connectors

External reference frequency input (EXT REF INPUT)

Frequency

1, 2, 5, and 10 MHz

 

(±200 Hz at 10 MHz)

Level

-10 dBm to +20 dBm, typical

Impedance

50 ohms

Connector

BNC (f)

High-stability frequency reference output (Option 1D5)

Frequency

10.0000 MHz

Frequency stability

±0.05 ppm

(0° to 55° C)

 

Daily aging rate

<3x10–9/day

(after 30 days)

 

Yearly aging rate

0.5 ppm/year

Output

0 dBm minimum

Nominal output impedance

50 Ω Connector BNC (f)

External auxiliary input (AUX INPUT)

Input voltage limits

-10V to +10V

External AM input (EXT AM)

±1 volt into a 5 k Ω resistor, 1 kHz maximum,

resulting in approximately 8 dB/volt amplitude modulation. BNC (f) connector.

External trigger (EXT TRIGGER)

Triggers on a negative TTL transition or contact closure to ground. BNC (f) connector.

20

Test sequence output (TEST SEQ)

By default, this connector outputs a TTL end-of- sweep signal. It can also be programmed by the user in a test sequence to output a user-defined TTL signal. BNC (f) connector.

Limit test output (LIMIT TEST)

This connector outputs a TTL signal of the limit test results. Pass: TTL high. Fail: TTL low. BNC (f) connector.

Test-port bias input (BIAS CONNECT) (8753ES only)

Maximum voltage

+30 VDC

Maximum current

±200 mA

(no degradation in RF specs)

 

Maximum current

±1 A

Connector

BNC (f)

VGA video output (EXT MON)

This connector drives external VGA monitors.

GPIB

This connector allows communications with compatible devices including external controllers, printers, plotters, disk drives, and power meters.

Parallel port

This 25-pin female connector is used with parallel peripherals. It can also be used as a general purpose I/O port.

RS-232C

This 9-pin male connector is used with serial peripherals.

DIN keyboard

This mini-DIN connector is used for adding an IBM PC-AT compatible keyboard.

Test set interconnect

This connector is used to connect an 8753ES Option 011 to the 85046A/B or 85047A test set. 8753ES analyzers without Option 011 can use signal levels on this connector for sequencing or general purpose I/O applications.

Internal memory

Typical data retention time with 3V, 1.2 Ah battery:

At 25° C

11,904 days (32.6 years)

At 40° C

1244 days (3.4 years)

At 70° C

250 days (0.68 year)

Line power

48 Hz to 66 Hz

115V nominal (90V to 132V) or 230V nominal (198V to 264V). 280 VA max.

Image 20
Contents Agilent 8753ET/ES 8753ET and 8753ES Network Analyzers, 30 kHz to 3 or 6 GHzTable of Contents Measurement Uncertainty Measurement Port CharacteristicsDynamic Range Configuration Agilent 8753ES 50-ohm systemsSystem dynamic range Reflection uncertaintyType-N test ports Calibration kit 85032B Test-port cables 11857DCalibration kit 85033D Test-port cables 11857D Calibration kit 85036B Test-port cables 11857B Agilent 8753ES 75-Ohm SystemsNetwork analyzer Type-F test ports Calibration kit 85039B Test-port cables 11857BMeasurement Port Characteristics Agilent 8753ET 50-ohm systemsPart number KHz to 16 MHz 100 dB MHz to 3 GHz 110 dBTest-Port Input Characteristics Agilent 8753ES SpecificationsTest-Port Output Characteristics7 External source mode1 CW time sweep only Phase CharacteristicsMagnitude Characteristics Test Port Input Characteristics Option Agilent 8753ES Option 011 SpecificationsTest Port Output Characteristics Channel input requirements1 Frequency offset mode3External source mode4 Agilent 8753ET Specifications Display resolution Input frequency accuracy requirement2Reference level Range Stability2 KHz to 3 GHzAgilent 8753ET/ES Supplemental Characteristics Group Delay CharacteristicsMeasurement Source ControlDisplay Control Data Accuracy EnhancementData Averaging Trace Functions StorageData Hardcopy System CapabilitiesRemote Programming Test SequencesTime-domain Option Rear-Panel Connectors General CharacteristicsFront Panel Connectors General Conditions Cabinet Dimensions 8753ET/ESOptions Agilent 85046A/B S-Parameter Test Sets Test Set Specifications for OptionAgilent 85046A/B Specifications Specifications Agilent 85047A S-Parameter Test SetAccessories Calibration KitsAgilent 85038 Family of Calibration Kits Agilent 11906 Family of Adapter KitsVerification Kits Test-Port Return Cables85024A High-Frequency Probe Ohm Accessory KitsRF Limiter Weight Net 3.5 kg 7.7 lb Agilent 8347A RF Amplifier
Related manuals
Manual 63 pages 28.88 Kb