Agilent Technologies manual Introduction to 82357A Operating Modes

Page 36

Using the 82357A

Modes of Operation

Introduction to 82357A Operating Modes

Single Mode Features

Multiple Mode Features

The 82357A has two modes of operation. When only one 82357A is connected to a USB port within a system, we define the operation as the single mode of operation. When two or more 82357As are connected at the same time to USB ports within a system, we define the operation as the multiple mode of operation.

NOTE

All SICL/VISA applications are notified when their 82357A has been removed from the system by returning VI_ERR_NOINFC (for VISA) or I_ERR_NCIC (for SICL).

For the single mode of operation, the operating parameters (VISA Interface ID, SICL Interface ID, Logical Unit Number, and GPIB Address) are set when the 82357A is first installed and configured.

If this 82357A is unplugged and replugged or if the 82357A is replaced with a different 82357A, the original configuration parameters are automatically assigned to the newly attached 82357A. Thus, you can exchange 82357As at any time without reconfiguring the interface. This allows exchanging 82357As among users, as long as only one 82357A is attached at any one time.

In contrast, when two or more 82357As are connected to the system at the same time, each 82357A must have its own specific set of operating parameters and each 82357A Serial Number is “bound” to its operating parameters. In multiple mode of operation, if you add a new 82357A or if you unplug an 82357A and plug in a new 82357A, the newly installed 82357A will be assigned a new (unique) set of operating parameters.

NOTE

Each time you attach a new 82357A, if Connection Expert is not running, an Agilent 82357A USB/GPIB Interface Detected dialog box will be displayed. If Connection Expert is running, the Connection Expert window will automatically refresh as the new 82357A is configured.

36

Chapter 2

Image 36
Contents USB/GPIB Interface Converter Page Contents Guide Contents Guide General Information Installing the 82357A Installing the 82357A Steps to Install the 82357A Before You Install the 82357ABefore You Install the 82357A Check Your ShipmentCheck System Requirements Installing Agilent IO Libraries Suite Checking for Installed Agilent IO LibrariesInstalling Agilent IO Libraries Suite ‰ Click Start Run Connecting the 82357A 82357A Hardware DescriptionConnecting the 82357A to Your PC Connecting the 82357A to a USB Hub USB Cable 82357A Observing Windows Plug and Play Manager Sequence Windows XP OnlyIntroduction Hardware Installation dialog box appears Typical Windows Plug and Play Manager SequenceConnecting the 82357A Connecting the 82357A Configuring the 82357A Setting 82357A Default ConfigurationConnecting Multiple 82357As. An Agilent 82357A USB/GPIB Configuring the 82357A Setting 82357A Custom Configuration Configuring the 82357A Connection Expert Connecting Gpib Instruments Connecting a Single Gpib InstrumentConnecting Multiple Gpib Instruments Programming via the 82357A Establishing Instrument CommunicationProgramming Gpib Instruments Accessing Visa Sicl Manuals ConfigurationIntroduction to IO InterfaceTypical System Installation 82357 USB/GPIB Interface Using the 82357A Using the 82357A Modes of Operation Initial 82357A Operating StatesSingle Mode Features Multiple Mode Features Introduction to 82357A Operating ModesSingle 82357A Operation Multiple 82357A Operation SRQ OperationSetting Configuration Parameters Changing Configuration ParametersSetting Timeout Floor Values Changing Modes of OperationSetting 82357A High-Performance Operation Introduction Setting T1 Value With VisaSetting T1 Value With Sicl Setting Configuration Parameters Chapter Troubleshooting the 82357A Troubleshooting the 82357A Troubleshooting Flowchart Observe the LED StatesFor telephone and web assistance, see Contacting Agilent on Reboot the PC Hardware ChecksCheck USB Cables, USB Interface, Host PC Check Device Manager Software Installation Checks Verify Agilent IO Libraries Suite InstallationCheck Suspend/Resume Operation Software Installation Checks Verify 82357A USB Driver Installation \Windows\inf\agt3579x.infSoftware Configuration Checks Checking IO Control OperationCheck USB Scanner Check for usbscan.sys Windows 98 SE Only Assigning usbscan Driver to the 82357A Service/Support Information 82357A Service InformationContacting Agilent 82357A User’s Guide Information 82357A User’s Guide Information Guide Contents Guide General Information Government Restricted RightsWarranty Information Trademark InformationPrinting History Copyright InformationDeclaration of Conformity 82357A Specifications 82357A Specifications Technical Specifications General RequirementsSupplementary Information Gpib Modes Operation SupportedIndex IndexIndex Page 82357-90002