Suunto VYTEC DS manual Oxygen Displays, Default Nitrox Settings

Page 33

on Nitrox use result in longer no-decompression times and shallower maximum depths than div- ing with air.

As a safety precaution the oxygen calculations in the computer are made with an oxygen per- centage of 1% + setO2% .

When the dive computer is set in Nitrox mode the Dive Planning and Dive Simulation modes both calculate with the O2% and PO2 values that are currently in the computer.

To set the Nitrox mixes refer to chapter 4.3 Set Modes.

DEFAULT NITROX SETTINGS

In Nitrox mode the Vytec allows the user to set 1-3 nitrox mixes containing 21-99% oxygen.

In the Nitrox mode, the default setting for MIX1 is standard air (21% O2). It remains in this setting until the O2% is adjusted to any other percentage of oxygen (22% - 100%). The default setting for maxi- mum oxygen partial pressure is 1.4 bar, however you are able to set it in the range of 0.5 - 1.6 bar.

If unused, the computer will retain the manually entered values for the oxygen percentage and maximum oxygen partial pressure for MIX1 for about two hours, after which it will revert to the default settings.

MIX2 and MIX3 are by default set to OFF. To set MIX2 and MIX 3 refer to chapter 3.4.4 Gas change and multiple breathing mixes. Oxygen percentages and maximum oxygen partial pres- sures for MIX2 and MIX3 are stored permanent- ly. If MIX2 and MIX3 are set OFF, the computer will retain the manually entered values for the oxygen percentage and maximum oxygen par- tial pressure for MIX1 for about two hours, after which it will revert to the default settings.

3.4.2. Oxygen Displays

If set to Nitrox mode the NITROX display, with all labeled oxygen information and the

 

m

CEILING

 

 

STOP ASC TIME

 

AIR TIME

 

C

 

 

B

 

 

T

bar

DIVE TIME

Fig. 3.19. Decompression dive, above ceiling.

Note downward pointing arrow, Er warning and an audible alarm. You should immediately (with- in 3 minutes) descend to or below ceiling.

CEILING

3m / 10ft

6m / 18ft

FLOOR

Fig. 3.20. Ceiling and Floor Zones. The Recom- mended and Manadatory Safety Stop zone lies between 6m and 3m [20 ft and 10ft].

m

PO2

O2%

OLF

Fig. 3.21. Nitrox display. Maximum depth based on set O2% (21%) and PO2 (1.4 bar) is 54.1 m [177 ft].

31

Image 33
Contents Suunto Vytec DS Guide Reference Quick COPYRIGHT, Trademark and Patent Notice Definition of WARNINGS, Cautions and NotesPrEN Page Page Page Table of Contents AIR/NITROX/GAUGE Suunto RGBM/ Deep Stop Algorithm IntroductionEmergency Ascents Safety PrecautionsDive Computer Limitations Getting AcquaintedNitrox FunctionsPush Buttons Making the Vytec personalWater Contacts Press the Smart Mode buttonInstalling the Transmitter Wireless Cylinder Pressure TransmissionPairing and Code Selection Transmission Data Setc Pressure Transmission Related DisplaysFail LobtDiving with the Vytec Before DivingActivation and Prechecks Battery Power Indicator Battery IndicationLowbat + Battery Power IndicatorWireless Transmitter Battery Indicator Dive Planning PlanDive Numbering Shown During Dive Planning User Definable Functions and Alarms Safety StopsRecommended Safety Stop Diving display. Pres- ent depth is 19.3 m Mandatory Safety StopDiving in AIR Mode Deep StopsBasic Dive Data Cylinder Pressure Data BookmarkConsumed Bottom Time CBT Ascent Rate Indicator Ascent Rate IndicatorDecompression dives Safety Stops and Deep StopsCEILING, Ceiling ZONE, Floor Decompression Range Display below the Floor Display Above the FloorDisplay AT the Ceiling Zone Display Above the Ceiling Diving in Nitrox Mode 3.4.1. Before DivingOxygen Displays Default Nitrox Settings22. Diving in Nitrox mode. The O2% is set to 32% Oxygen Limit Fraction OLFGas Change, Multiple Breathing Gas Mixes Diving in Gauge Mode AT the Surface 3.6.1. Surface Interval Separated by a colon, telling Dive NumberingFlying After Diving Three double beeps occur, when Audible and Visual AlarmsSet oxygen partial pressure limit is exceeded. The maximum Oxygen Alarms in Nitrox ModeAltitude Adjustment High Altitude Dives and Personal AdjustmentPersonal Adjustment Altitude Adjustment RangesPersonal Adjustment Ranges Error Conditions Rgbm Model SettingsOmitted Decompression USE of the Menu Based Functions Menu Based ModesList of the Menu Based Modes Exit / QuitLogbook and Dive Profile Memory 1 Logbook Memories and Data Transfer 1 MemoryI, main display Fig III Fig Dive Profile Memory Prof Dive History Memory 2 HistoryData Transfer and PC-Interface 3 TR-PC Dive Simulator 1 Simdive Simulation Mode 2 SimulSET Modes 3 SET Dive Planning Simulator 2 SimplanDive Parameter Settings 1 SET Dive Dive Time Alarm Setting 2 d Alarm Maximum Depth Alarm Setting 3 MAX DpthNitrox/Oxygen Settings 4 Nitrox Adjusting Time 1 AdJ Time Setting Time 2 SET TimeAdjusting Daily Alarm 3 T Alarm Adjusting Date 2 AdJ DateSetting Preferences 3 SET Pref Backlight Setting 1 LightWireless Transmission Settings 3 HP Dive Computer Units Setting 2 UnitsProfile Sample Rate Setting 4 REC Dive Computer Model Setting 5 ModelImportant Information Care and MaintenanceCare of Your Dive Computer Water Contacts and Push ButtonsWater Resistance Inspection MaintenanceBattery KIT Battery Replacement 5.5.1. Computer batteryTools Required Battery ReplacementPage Opening of the securing ring Wireless Transmitter Battery ReplacementTools Required Transmitter Battery ReplacementTechnical Description Operating PrinciplesNO-DECOMPRESSION Limits Altitude Diving Surface Intervals Suunto Reduced Gradient Bubble Model RgbmOxygen Exposure Transmitter Technical SpecificationOther displays Battery Vytec Warranty System Requirements SUUNTOSPORTS.COMCommunities My SuuntoGetting Started Sport ForumsGlossary EAN EADNoaa NitroxNo DEC Time OLFPO2 RgbmSurf Time Tissue group Disposal of the Device TIME-KEEPING and STAND-BY Mode Page Page Made in Finland Suunto Oy 01 / 2004, 08