Suunto VYTEC DS manual Dive Numbering, Separated by a colon, telling

Page 38

 

 

Or when the TIME button is pressed once or

 

m

twice:

 

 

 

the current time, shown as TIME

 

MAX

 

 

 

 

instead of the DIVE TIME

NO

SURF

 

 

 

TIME

 

the current temperature with °C for

°C

TIME

 

 

Centigrade [or °F for Fahrenheit]

 

 

 

the surface time in hours and minutes

 

 

 

 

(separated by a colon), telling the

Fig 3.30. Surface interval, Sur-

 

duration of the present surface interval

 

(Fig. 3.30.)

face time display. Pressing TIME

 

the desaturation/no-flying time in

button once will show surface

 

time display.

 

 

hours and minutes is shown next to

 

 

 

 

the airplane in the center window of

 

 

 

 

the display (Fig. 3.31.).

 

m

If set to Nitrox mode, the following information

 

MAX

will also be shown:

NO

 

 

the oxygen percentage labeled with

 

 

 

 

 

 

 

O2% is shown on the left side of the

°C

TIME

 

 

center window

 

 

 

the current oxygen toxicity exposure

 

 

 

 

shown with an Oxygen Limit Fraction

Fig. 3.31. Surface interval,

 

 

(OLF) bar graph along the left side of

 

 

the display.

no-flying time. Pressing

 

 

TIME button twice will

3.6.2. Dive Numbering

show no-flying time,

indicated by an airplane

Several repetitive dives are considered to be-

symbol.

 

 

 

long to the same repetitive dive series when

 

 

the dive computer has not counted the no-fly-

 

m

ing time to zero. Within each series, the dives

 

MAX

are given individual numbers. The first dive of

NO

 

the series will be numbered as DIVE 1, the

second as DIVE 2, the third as DIVE 3, etc.

°C

DIVE TIME

Fig. 3.32. Surface Mode after a violated decom-pres- sion dive. Er symbol indi- cates that you have violated the ceiling for

more than three minutes. You must not dive again for at least 48 hours.

If you start a new dive with less than 5 minutes of surface interval time, the dive computer in- terprets this as a continuation of the previous dive and the dives are considered to be the same. The diving display will return, the dive number will remain unchanged, and the dive time will begin where it left off. After 5 minutes on the surface, subsequent dives are, by defi- nition, repetitive. The dive counter displayed

36

Image 38
Contents Suunto Vytec DS Guide Reference Quick Definition of WARNINGS, Cautions and Notes COPYRIGHT, Trademark and Patent NoticePrEN Page Page Page Table of Contents AIR/NITROX/GAUGE Introduction Suunto RGBM/ Deep Stop AlgorithmSafety Precautions Emergency AscentsNitrox Getting AcquaintedDive Computer Limitations FunctionsMaking the Vytec personal Push ButtonsPress the Smart Mode button Water ContactsWireless Cylinder Pressure Transmission Installing the TransmitterPairing and Code Selection Transmission Data Fail Pressure Transmission Related DisplaysSetc LobtActivation and Prechecks Diving with the VytecBefore Diving Battery Indication Battery Power IndicatorBattery Power Indicator Lowbat +Dive Numbering Shown During Dive Planning Wireless Transmitter Battery IndicatorDive Planning Plan Recommended Safety Stop User Definable Functions and AlarmsSafety Stops Mandatory Safety Stop Diving display. Pres- ent depth is 19.3 mBasic Dive Data Diving in AIR ModeDeep Stops Bookmark Cylinder Pressure DataConsumed Bottom Time CBT Ascent Rate Indicator Ascent Rate IndicatorSafety Stops and Deep Stops Decompression divesCEILING, Ceiling ZONE, Floor Decompression Range Display AT the Ceiling Zone Display below the FloorDisplay Above the Floor Diving in Nitrox Mode 3.4.1. Before Diving Display Above the CeilingDefault Nitrox Settings Oxygen Displays22. Diving in Nitrox mode. The O2% is set to 32% Gas Change, Multiple Breathing Gas Mixes Oxygen Limit FractionOLF Diving in Gauge Mode AT the Surface 3.6.1. Surface Interval Dive Numbering Separated by a colon, tellingFlying After Diving Audible and Visual Alarms Three double beeps occur, whenOxygen Alarms in Nitrox Mode Set oxygen partial pressure limit is exceeded. The maximumPersonal Adjustment High Altitude Dives and Personal AdjustmentAltitude Adjustment Altitude Adjustment RangesPersonal Adjustment Ranges Omitted Decompression Error ConditionsRgbm Model Settings Menu Based Modes USE of the Menu Based FunctionsExit / Quit List of the Menu Based ModesMemories and Data Transfer 1 Memory Logbook and Dive Profile Memory 1 LogbookI, main display Fig III Fig Dive History Memory 2 History Dive Profile Memory ProfData Transfer and PC-Interface 3 TR-PC Simulation Mode 2 Simul Dive Simulator 1 SimdiveDive Planning Simulator 2 Simplan SET Modes 3 SETDive Parameter Settings 1 SET Dive Nitrox/Oxygen Settings 4 Nitrox Dive Time Alarm Setting 2 d AlarmMaximum Depth Alarm Setting 3 MAX Dpth Setting Time 2 SET Time Adjusting Time 1 AdJ TimeSetting Preferences 3 SET Pref Adjusting Date 2 AdJ DateAdjusting Daily Alarm 3 T Alarm Backlight Setting 1 LightProfile Sample Rate Setting 4 REC Dive Computer Units Setting 2 UnitsWireless Transmission Settings 3 HP Dive Computer Model Setting 5 ModelCare of Your Dive Computer Care and MaintenanceImportant Information Water Contacts and Push ButtonsMaintenance Water Resistance InspectionBattery Replacement 5.5.1. Computer battery Battery KITBattery Replacement Tools RequiredPage Wireless Transmitter Battery Replacement Opening of the securing ringTransmitter Battery Replacement Tools RequiredNO-DECOMPRESSION Limits Technical DescriptionOperating Principles Altitude Diving Suunto Reduced Gradient Bubble Model Rgbm Surface IntervalsOxygen Exposure Technical Specification TransmitterOther displays Battery Vytec Warranty SUUNTOSPORTS.COM System RequirementsMy Suunto CommunitiesSport Forums Getting StartedGlossary EAD EANNo DEC Time NitroxNoaa OLFSurf Time PO2Rgbm Tissue group Disposal of the Device TIME-KEEPING and STAND-BY Mode Page Page Made in Finland Suunto Oy 01 / 2004, 08