Emerson MAS 2600 user manual Deck Box 165B9035

Page 15

 

Installation & User Manual

June 2013

Ed 19UK

 

 

Amplifier PCB MAS2600-XXX-XX-X/XX

The PCB is available in different types for the different transducers, with or without terminals for Pt100 temperature sensor.

0:Without amplifier PCB.

P or S: Programmable output range. The amplifier can be calibrated to 4-20 mA output in whatever range wanted within the transducer range. Calibration is made by means of dip switches and potentiometers.

D or M: Differential output range. The output can be adjusted to 12 mA output with no pressure applied. With pressure or vacuum, the output goes against 20 mA and 4 mA respectively. Full-scale and zero point calibration by means of software, e.g. tank monitoring software.

The types S and M are with terminals for temperature sensors.

The types P and D are without terminals for temperature sensors.

Programmable Range (Type P and S)

Deck Box 165B9035

Page 15 of 28

Image 15
Contents MAS2600 Table of contents MAS2600 Tank Contents Transmitter Description and OperationApplication Technical Specifications Transducer Ranges MAS2600 Ordering InformationGauge Transmitter TypeTransducer type 3.5 bar abs 2.0 bar abs Gauge and High Temperature Gauge TransducersSelecting the Measuring Range Selecting the Transducer TypeAbsolute Transducers Basic RulesForepeak Tank Draft Temperature Sensor MAS2600-XXX-XX-X/XXPole Mounting drawing no. G022P011 Brackets for Internal Mounting drawing no. G022P010Flange Mounting drawing no. G022P013 Pipe End Mounting drawing no. G022P015Flexible rubber tube mounting 80 ºC drawing no. G022P001 DN25 flange with 1 ball valve drawing no. G022P019 75.0 Deck Box 165B9035 Handling Installation June XXXXXX-YYYY Intrinsically Safe InstallationEd 19UK Any 28V / 300R Testing Testing and Recalibrating= 12 mA + P Trd .range −Calculated output at atmospheric pressure Maintenance and Trouble Shooting MaintenanceTrouble Shooting FAX MAS2600 Test sheetEmerson Process Management