Emerson Comprehensive Installation and User Guide for MAS Series Marine Instruments

Page 19

 

Installation & User Manual

June 2013

Ed 19UK

 

 

Intrinsically Safe Installation

The MAS2600 is designed for use in an approved intrinsically safe installation. The amplifier is fitted with a label describing the IS category:

S/N: XXXXXX-YYYY

DEMKO 11 ATEX 146506X IECEx ULD 10.0013X

Ui: 29V Ii: 93 mA Li: 20µH Ci: 60 nF Pi: 0.68 W

II 2(1) G

Ex ia IIC T4 Tamb.: -20 to +80 °C

Damcos A/S, Aaderupvej 41,DK-4700, Naestved

Where

 

 

S/N

:

serial number – production year.

 

:

Ex identification according to guideline 94/9/EC (ATEX).

II

:

other aria than mines.

2(1)

:

sensor mounted in zone 0 and amplifier in zone 1

Ex ia

:

denotes the type of protection intrinsic safety , i.e. the system is incapable of causing

 

 

ignition in normal operation or with any combination of up to two faults applied. This

 

 

category will be adequate for use in zone 0, 1 and 2.

IIC

:

denotes the group of gases.

T4

:

denotes the maximum surface temperature classification specified (T4: 135 °C).

Tamb

:

denotes the minimum and maximum ambient temperature for the amplifier box

permitted by the certificate.

DEMKO 11 ATEX 146506X:

The ATEX Certificate No 146506 is issued by DEMKO 201 with remarks. Inmetro:

Inmetro Certificate No. UL BR 12.0116X issued by UL do Brazil

Relevant intrinsic safety certificates to be supplied with the system units.

The certification of electric apparatus as intrinsically safe requires the apparatus to be designed and tested in accordance with approved drawings and the apparatus to be installed in accordance with approved block schematic diagrams.

Page 19 of 28

Image 19
Contents MAS2600 Table of contents Description and Operation MAS2600 Tank Contents TransmitterApplication Technical Specifications Transducer Ranges MAS2600 Ordering InformationGauge Transmitter TypeTransducer type 3.5 bar abs 2.0 bar abs Gauge and High Temperature Gauge TransducersSelecting the Measuring Range Selecting the Transducer TypeBasic Rules Absolute TransducersForepeak Tank Draft Temperature Sensor MAS2600-XXX-XX-X/XXPole Mounting drawing no. G022P011 Brackets for Internal Mounting drawing no. G022P010Flange Mounting drawing no. G022P013 Pipe End Mounting drawing no. G022P015Flexible rubber tube mounting 80 ºC drawing no. G022P001 DN25 flange with 1 ball valve drawing no. G022P019 75.0 Deck Box 165B9035 Handling Installation June XXXXXX-YYYY Intrinsically Safe InstallationEd 19UK Any 28V / 300R Testing Testing and Recalibrating= 12 mA + P Trd .range −Calculated output at atmospheric pressure Maintenance Maintenance and Trouble ShootingTrouble Shooting FAX MAS2600 Test sheetEmerson Process Management