Recalibrating the Absolute Transmitter using Test Cup type G022S100
When recalibrating the MAS2600 Absolute Transmitter a pressure calibrator with a range of
Atmospheric pressure and Transducer range to be entered into the formulas as mbar.
1. | The calibrator with the output selector switch set to pressure mode is connected to the |
| MAS2600 transmitter via a Test Cup, type G022S100. |
2.Check the transducer range.
The transducer range is indicated on the lid inside the amplifier box.
3.Calculate output signal in conditions of no pressure (atmospheric pressure), using below formula. The output signal should be adjusted to the calculated value on the zero potentiometer.
mA @ Atmospheric pressure = 4 + | 16*( Atm. press − 800) | |
Trd.range − 800 | ||
|
4.Calculate the pressure for maximum output signal, using the following formula. mbarg @ 20 mA = Trd.range − Atm. press
Increase the pressure to calculated pressure and adjust the output signal to 20.00 mA on the span potentiometer.
5.Remove the pressure and check the zero point.
6.Check the linearity at
Output signal at different pressures P can be calculated, using the following formula.
mA @ P mbarg = 4 + | 16*( Atm. press − 800 | + P) |
Trd.range − 800 |
| |
|
|
7.If necessary, repeat the steps 3 to 6.