Recalibrating the Absolute Transmitter using Test Cup type G022S100

When recalibrating the MAS2600 Absolute Transmitter a pressure calibrator with a range of -1 to 3.5 bar should be used. It should have an accuracy of 0.05% FS or better, as well as a 4½ digit milliampmeter. If a calibrator without reference transmitter is used, the atmospheric pressure must be known to calculate the correction.

Atmospheric pressure and Transducer range to be entered into the formulas as mbar.

1.

The calibrator with the output selector switch set to pressure mode is connected to the

 

MAS2600 transmitter via a Test Cup, type G022S100.

2.Check the transducer range.

The transducer range is indicated on the lid inside the amplifier box.

3.Calculate output signal in conditions of no pressure (atmospheric pressure), using below formula. The output signal should be adjusted to the calculated value on the zero potentiometer.

mA @ Atmospheric pressure = 4 +

16*( Atm. press 800)

Trd.range 800

 

4.Calculate the pressure for maximum output signal, using the following formula. mbarg @ 20 mA = Trd.range Atm. press

Increase the pressure to calculated pressure and adjust the output signal to 20.00 mA on the span potentiometer.

5.Remove the pressure and check the zero point.

6.Check the linearity at 0-25-50-75 and 100% of the measuring range.

Output signal at different pressures P can be calculated, using the following formula.

mA @ P mbarg = 4 +

16*( Atm. press 800

+ P)

Trd.range 800

 

 

 

7.If necessary, repeat the steps 3 to 6.

Page 24 of 28

Image 24
Contents MAS2600 Table of contents MAS2600 Tank Contents Transmitter Description and OperationApplication Technical Specifications MAS2600 Ordering Information Transducer RangesType Gauge TransmitterGauge and High Temperature Gauge Transducers Selecting the Measuring RangeSelecting the Transducer Type Transducer type 3.5 bar abs 2.0 bar absAbsolute Transducers Basic RulesForepeak Tank Temperature Sensor MAS2600-XXX-XX-X/XX DraftBrackets for Internal Mounting drawing no. G022P010 Pole Mounting drawing no. G022P011Pipe End Mounting drawing no. G022P015 Flange Mounting drawing no. G022P013Flexible rubber tube mounting 80 ºC drawing no. G022P001 DN25 flange with 1 ball valve drawing no. G022P019 75.0 Deck Box 165B9035 Handling Installation June Intrinsically Safe Installation XXXXXX-YYYYEd 19UK Any 28V / 300R Testing and Recalibrating Testing= 12 mA Trd .range − + PCalculated output at atmospheric pressure Maintenance and Trouble Shooting MaintenanceTrouble Shooting MAS2600 Test sheet FAXEmerson Process Management