Emerson MAS 2600 user manual Maintenance and Trouble Shooting

Page 26

Installation & User Manual

Ed 19UK

June 2013

 

 

Maintenance and Trouble Shooting

Maintenance

Damcos recommends that the following to be checked annually:

1.Visual inspection of the cable, transducer and amplifier.

2.Compare signal with e.g. hand sounding, and if necessary perform a zero and full-scale calibration.

Trouble Shooting

If the MAS2600 transmitter is not functioning in accordance with Emerson Process Management Marine Solutions, Damcos instructions, check that:

1.the 24 VDC power supply has been correctly connected, and is between 14-33 VDC.

2.the 4 - 20 mA output signal is functioning (compared with e.g. hand sounding).

3.the transducer cable has been correctly fitted and the shield connected to the shield terminal in the amplifier box.

4.the diaphragm has not been damaged.

If this does not succeed, fill in the test sheet on page 27 and submit this to our Service department fax no.: +45 5578 7272.

When contacting our Service department please specify the serial numbers of both transducer and amplifier.

The test sheet will be examined to see whether it can be established that the transmitter is broken or not.

When returning a defective MAS2600 transmitter for repairs, please ensure that:

The transducer diaphragm is protected against blows of any kind.

Both the transducer and the amplifier are returned.

A detailed description of the fault is enclosed.

Page 26 of 28

Image 26
Contents MAS2600 Table of contents Application MAS2600 Tank Contents TransmitterDescription and Operation Technical Specifications MAS2600 Ordering Information Transducer RangesType Gauge TransmitterSelecting the Transducer Type Gauge and High Temperature Gauge TransducersSelecting the Measuring Range Transducer type 3.5 bar abs 2.0 bar absForepeak Tank Absolute TransducersBasic Rules Temperature Sensor MAS2600-XXX-XX-X/XX DraftBrackets for Internal Mounting drawing no. G022P010 Pole Mounting drawing no. G022P011Pipe End Mounting drawing no. G022P015 Flange Mounting drawing no. G022P013Flexible rubber tube mounting 80 ºC drawing no. G022P001 DN25 flange with 1 ball valve drawing no. G022P019 75.0 Deck Box 165B9035 Handling Installation June Intrinsically Safe Installation XXXXXX-YYYYEd 19UK Any 28V / 300R Testing and Recalibrating Testing= 12 mA Trd .range − + PCalculated output at atmospheric pressure Trouble Shooting Maintenance and Trouble ShootingMaintenance MAS2600 Test sheet FAXEmerson Process Management