Teledyne 465L operation manual Principle of Operation, = − InHg × ln ×  273

Page 21

3.PRINCIPLE OF OPERATION

The detection of ozone molecules is based on absorption of 254 nm UV light due to an internal electronic resonance of the O3 molecule. The Model 465 uses a mercury lamp constructed so that a large majority of the light emitted is at the 254nm wavelength. Light from the lamp shines down a hollow quartz tube that is alternately filled with sample gas, then filled with gas scrubbed to remove ozone. The ratio of the intensity of light passing through the scrubbed gas to that of the sample forms a ratio I/Io. This ratio forms the basis for the calculation of the ozone concentration.

The Beer-Lambert equation, shown below, calculates the concentration of ozone from the ratio of light intensities.

C = −

109

×

Τ

 

×

29.92inHg

× ln

Ι

 

o

 

 

 

O3

α ×

Κ

 

Ρ

 

Ιo

 

273

 

 

Where:

I = Intensity of light passed through the sample

Io = Intensity of light through sample free of ozone

α= absorption coefficient = path length

CO3 = concentration of ozone in ppb

T= sample temperature in degrees Kelvin P = pressure in inches of mercury

As can be seen the concentration of ozone depends on more than the intensity ratio. Temperature and pressure influence the density of the sample. The density changes the number of ozone molecules in the absorption tube which impacts the amount of light removed from the light beam. These effects are addressed by directly measuring temperature and pressure and including their actual values in the calculation. The absorption coefficient is a number that reflects the inherent ability of ozone to absorb 254 nm light. Most current measurements place this value at 308 cm-1atm-1at STP. The value of this number reflects the fact that ozone is a very efficient absorber of UV radiation which is why stratospheric ozone protects the life forms lower in the atmosphere from the harmful effects from solar UV radiation. Lastly, the absorption path length determines how many molecules are present in the column of gas in the absorption tube.

05509F DCN6409

19

Image 21
Contents Model 465L Ozone Monitor Page About Teledyne Advanced Pollution Instrumentation Tapi This page intentionally left blank General Safety Hazard Important Safety InformationMise EN Garde Consignes DE SécuritéPart Name/Description Number About this ManualRevision History Table of Contents Digital Communications Figures Sensor Detail Tables Xii 05509F DCN6409 Preface IntroductionThis page intentionally left blank Specifications Specifications and Agency ApprovalsMulti-Stream Specifications SafetyApprovals EMCPhone Other Type CertificationsThis page intentionally left blank = − InHg × ln ×  273 Principle of Operation05509F DCN6409 Getting Started Nema AC Power Connection to Monitor Pneumatic Connections Pneumatic Connections, Rack Mount ConfigurationSample Gas Connection Pneumatic Connections, Nema Configuration, Multi-StreamSample System Design in Humid Environments Sample Filtration05509F DCN6409 Permeation Gas Dryer Option Sample Conditioner OptionRear Panel Electrical I/O Connections Exhaust ConnectionElectrical I/O Connections Location of I/O Connectors in the Nema ConfigurationAnalog Output Status OutputsStatus Name Normal Operating Output # State Status Output SchematicRelay # Function Normal Operating State Hi-Current Relay OutputsRelay # Function Normal Operating StateRelay # Function Meaning Multi-Stream Relay Outputs Optional6.1. RS232 Connection Digital Serial Interfaces6.2. RS485 Connection Ethernet ConnectionHis page intentionally left blank Frequently Asked Questions This page intentionally left blank Options 05509F DCN6409 Front Panel Overview OperationOzone Concentration Display Menu Navigation KeypadConcentration Alarm Status/Reset Alarm Status LED Color Alarm StateStream Selector Status/Control Optional Front Panel Menus Status LEDsMenu Navigation Basics Status LED Normal StateEditing Numeric Values Front Panel Menu Diagram OFF Diag Menu Analog Step-TestPoint 5V Output MA Output Signal Name Rear Panel Label Function Signal I/OZero Calibration Memory ResetSpan Calibration Span Cal Menu Flow Calibration Optional Electronic Flowmeter onlyFlow Cal Menu View Menu Config MenuParameter Meaning UnitsCount SETUP-VARS MenuProperty Meaning Default Range Value SETUP-STREAM MenuMulti-Stream Operation Changing Com Port Hardware Configuration Digital CommunicationsDigital Communications Overview RS232 Configuration RS-232 Configuration Function DB-9 General I/O Connector Pin SW2 DTE SW2 DCEProperty Value Function DB-9 General I/O Connector Pin SW2 DCE RS-485 ConfigurationHardware Setup for Configuring the Ethernet Module Ethernet ConfigurationVerifying Network Hardware Connection Link LED Left Side Activity LED Right Side Color MeaningManually Assigning a Static IP Address Determining a Dynamically Assigned IP Address Dhcp ModeModbus Register Maps Modbus Commands SupportedDescription Function CodeDecimal Address Description Register Type Start Bit # Bits Real-Time Concentration and Status Registers Read OnlyInstrument Setup and Configuration Registers Read / Write Decimal Address Description Register Type Start Bit # Bits This page intentionally left blank Maintenance and Adjustments Maintenance ScheduleMaintenance Item Recommended Interval Section Instrument Layout Instrument Layout, Rack Mount ConfigurationReplacing Internal Particulate Filter Instrument Layout, Nema ConfigurationUV Lamp Adjustment Sample Pump Replacement UV Lamp Replacement05509F DCN6409 Sensor Detail Sensor Valve ReplacementConfiguring the Analog Output Water Drop-Out Filter Option MaintenanceBoot Loader Operation Cleaning Exterior Surfaces of the M465LBoot Loader Hardware Setup Boot Loader Software SetupUpload Firmware Verify Firmware This page intentionally left blank Troubleshooting Pneumatic DiagramReference Drawings RACK-MOUNT Version only Interconnect DiagramStatus LED/Output Definitions Sensor OK Pneumatic ErrorInvalid Reading Check LampPressure Too Low Pressure Too HighThis page intentionally left blank Calibration Factory CalibrationZero Calibration This page intentionally left blank Spare Parts 12.1. M465L Rack Mount Spare Parts ListPart Number Description 12.2. M465L Nema Spare Parts ListCONNECTOR, 16 PIN, W/SCREW Flange This page intentionally left blank Permeation Gas Dryer Index