Teledyne 465L Stream Selector Status/Control Optional, Concentration Alarm Status/Reset

Page 42

Teledyne API Model 465L O3 Monitor Operation Manual

7.1.3.Stream Selector Status/Control (Optional)

The optional multi-channel stream selector can be controlled via the front panel. The stream selector operates in two modes, Auto and Manual. The Auto mode is normally the default mode. The ‘AUTO’ and ‘MAN’ LED’s show the current mode. To switch modes, press and hold the SELECT button for at least 2 seconds. When in Manual mode, momentarily pressing the SELECT button will switch to the next stream.

The stream selector status is shown by the six LED’s labeled ‘SAMPLE POINT.’ One of the LED’s should always be lit, showing the current channel being measured. Note that for a three- channel instrument, only streams 1-3 can be measured. See Section 7.3 for additional details on the operation of the stream selector.

When the instrument is in its default configuration, there is a Manual mode timeout feature that will automatically return the stream selector to Auto mode when no keys are pressed for 10 minutes. This is the recommended setting for instruments used for critical safety applications. This feature prevents the instrument from being inadvertently left in Manual mode, and subsequent loss of monitoring data from the other, unmeasured channels. For applications where Manual mode is the preferred mode of operation, this timeout feature can be turned off. See Section 7.2.6 for details on changing this setting.

7.1.4.Concentration Alarm Status/Reset

The M465L front panel displays the concentration alarm status of all channels being sampled. The row of LED’s labeled ‘ALARM STATUS’ shows the alarm status for each channel as follows:

Table 7-1: Alarm LED’s

Alarm Status LED Color

Alarm State

Green

No alarm active

Amber

‘HI’ alarm active

Red

‘HI-HI’ alarm active

NOTE

If the alarms are configured for latching operation, then the RESET button can be used to attempt to reset all alarms. Note that alarms can only be reset if the concentration measured for a specific channel has subsequently dropped below the alarm limit. If the alarms are configured for non-latching operation, they will be automatically cleared when the concentration falls below the alarm threshold. The RESET button has no effect when the alarms are configured for non- latching operation.

See Sections 7.2.7, 4.2.4, and 4.2.5 for additional details on concentration alarms.

40

05509F DCN6409

Image 42
Contents Model 465L Ozone Monitor Page About Teledyne Advanced Pollution Instrumentation Tapi This page intentionally left blank Important Safety Information General Safety HazardConsignes DE Sécurité Mise EN GardeAbout this Manual Part Name/Description NumberRevision History Table of Contents Digital Communications Figures Sensor Detail Tables Xii 05509F DCN6409 Introduction PrefaceThis page intentionally left blank Specifications and Agency Approvals SpecificationsApprovals SafetyMulti-Stream Specifications EMCOther Type Certifications PhoneThis page intentionally left blank Principle of Operation = − InHg × ln ×  27305509F DCN6409 Getting Started Nema AC Power Connection to Monitor Pneumatic Connections, Rack Mount Configuration Pneumatic ConnectionsPneumatic Connections, Nema Configuration, Multi-Stream Sample Gas ConnectionSample Filtration Sample System Design in Humid Environments05509F DCN6409 Sample Conditioner Option Permeation Gas Dryer OptionExhaust Connection Rear Panel Electrical I/O ConnectionsAnalog Output Location of I/O Connectors in the Nema ConfigurationElectrical I/O Connections Status OutputsStatus Output Schematic Status Name Normal Operating Output # StateRelay # Function Hi-Current Relay OutputsRelay # Function Normal Operating State Normal Operating StateMulti-Stream Relay Outputs Optional Relay # Function Meaning6.2. RS485 Connection Digital Serial Interfaces6.1. RS232 Connection Ethernet ConnectionHis page intentionally left blank Frequently Asked Questions This page intentionally left blank Options 05509F DCN6409 Ozone Concentration Display OperationFront Panel Overview Menu Navigation KeypadConcentration Alarm Status/Reset Alarm Status LED Color Alarm StateStream Selector Status/Control Optional Menu Navigation Basics Status LEDsFront Panel Menus Status LED Normal StateEditing Numeric Values Front Panel Menu Diagram OFF Diag Menu Analog Step-TestPoint 5V Output MA Output Signal I/O Signal Name Rear Panel Label FunctionMemory Reset Zero CalibrationSpan Calibration Flow Calibration Optional Electronic Flowmeter only Span Cal MenuFlow Cal Menu Parameter Meaning Config MenuView Menu UnitsSETUP-VARS Menu CountSETUP-STREAM Menu Property Meaning Default Range ValueMulti-Stream Operation Changing Com Port Hardware Configuration Digital CommunicationsDigital Communications Overview RS232 Configuration RS-232 Configuration Function DB-9 General I/O Connector Pin SW2 DTE SW2 DCEProperty Value RS-485 Configuration Function DB-9 General I/O Connector Pin SW2 DCEVerifying Network Hardware Connection Ethernet ConfigurationHardware Setup for Configuring the Ethernet Module Link LED Left Side Activity LED Right Side Color MeaningDetermining a Dynamically Assigned IP Address Dhcp Mode Manually Assigning a Static IP AddressDescription Modbus Commands SupportedModbus Register Maps Function CodeReal-Time Concentration and Status Registers Read Only Decimal Address Description Register Type Start Bit # BitsInstrument Setup and Configuration Registers Read / Write Decimal Address Description Register Type Start Bit # Bits This page intentionally left blank Maintenance and Adjustments Maintenance ScheduleMaintenance Item Recommended Interval Section Instrument Layout, Rack Mount Configuration Instrument LayoutInstrument Layout, Nema Configuration Replacing Internal Particulate FilterUV Lamp Adjustment UV Lamp Replacement Sample Pump Replacement05509F DCN6409 Sensor Valve Replacement Sensor DetailWater Drop-Out Filter Option Maintenance Configuring the Analog OutputCleaning Exterior Surfaces of the M465L Boot Loader OperationBoot Loader Hardware Setup Boot Loader Software SetupUpload Firmware Verify Firmware This page intentionally left blank Troubleshooting Pneumatic DiagramReference Drawings Interconnect Diagram RACK-MOUNT Version onlyStatus LED/Output Definitions Invalid Reading Pneumatic ErrorSensor OK Check LampPressure Too High Pressure Too LowThis page intentionally left blank Calibration Factory CalibrationZero Calibration This page intentionally left blank 12.1. M465L Rack Mount Spare Parts List Spare Parts12.2. M465L Nema Spare Parts List Part Number DescriptionCONNECTOR, 16 PIN, W/SCREW Flange This page intentionally left blank Index Permeation Gas Dryer